MalfunctionCKA

ALANINE AMINOTRANSFERASE

Received Nov 21, 2025 · Event occurred Oct 11, 2025

Report 1823260-2025-04769 · MDR key 23618540

Device

Generic name

Nadh Oxidation/nad Reduction, Alt/sgpt

Manufacturer

Roche Diagnostics

Catalog number

ASKU

Lot number

ASKU

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Output Problem
  • Incorrect, Inadequate or Imprecise Result or Readings
  • Output Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

BASED ON THE INFORMATION PROVIDED, THE INVESTIGATION DETERMINED THAT THE EVENT WAS CONSISTENT WITH THE SAMPLE CONTAINING A SUBSTANCE THAT STRONGLY ABSORBS AT 340 NM, THE MEASURING WAVELENGTH OF THE ASSAY. THE EVENT WAS A SAMPLE-SPECIFIC ISSUE. THERE WAS NO PRODUCT PROBLEM IDENTIFIED.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE RESULTS FROM THE COBAS 8000 C702 MODULE. THE CUSTOMER SUSPECTED INTERFERENCE, AS THE SAMPLE IN QUESTION WAS COLLECTED 17 HOURS AFTER THE PATIENT HAD RECEIVED RED BLOOD CELLS DUE TO BEING ANEMIC. THE INITIAL AST (ASPARTATE AMINOTRANSFERASE) RESULT WAS 17.6 U/L WITH AN ANALYZER ALARM. THE INITIAL ALT (ALANINE AMINOTRANSFERASE) RESULT WAS 11.3 U/L WITH AN ANALYZER ALARM. DUE TO THE ALARM, A 1:10 AUTO DILUTION WAS PERFORMED. THE RESULT FOR AST WAS 2.2 U/L, AND THE RESULT FOR ALT WAS 10.3 U/L. A NEW SAMPLE WAS COLLECTED FROM THE PATIENT ON (B)(6) 2025, AND THE AST AND ALT RESULTS HAD NO FLAGS. NO SPECIFIC DATA WAS PROVIDED. THIS MEDWATCH IS FOR ALT (ALANINE AMINOTRANSFERASE). REFER TO THE MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR THE AST (ASPARTATE AMINOTRANSFERASE) ASSAY.

Additional Manufacturer Narrative

THE COBAS 8000 C702 MODULE SERIAL NUMBER WAS (B)(6). THE INVESTIGATION IS ONGOING.

Additional Manufacturer Narrative

ON 19-DEC-2025, FURTHER TESTING OF THE SAMPLE GENERATED THE FOLLOWING RESULTS: ALT 6.8 U/L WITH A DATA FLAG AND 7.8 U/L WITH A DATA FLAG. AST 16.7 U/L WITH A DATA FLAG AND 12.9 U/L. THE INVESTIGATION IS ONGOING.