Received Nov 17, 2025 · Event occurred Jun 1, 2025
Report 1416980-2025-05699 · MDR key 23576825
Device
Generic name
Filter, Infusion Line
Manufacturer
Baxter Healthcare CorporationModel number
2N3350Catalog number
2N3350
Lot number
DR24J15017
Product problems
- Break
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WAS ADDED TO D9, H3, H4, H6, AND H11: H11: THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. A PRESSURE TEST WAS PERFORMED, AND NO LEAKS WERE OBSERVED. THE REPORTED CONDITION WAS NOT VERIFIED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative
SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THAT THE MEDIPORT NEEDLE OF A MICRO-VOLUME EXTENSION SET BROKE DURING PATIENT INFUSION. THE DEVICE CONTAINED IVF AND MESNA. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.