CYSTO-URETHRO-FIBERSCOPE

Received Nov 14, 2025

Report 1221826-2025-02005 · MDR key 23553697

Device

Generic name

Cysto-urethro-fiberscope

Model number

11272C1

Catalog number

11272C1

Product problems

  • Fracture
  • Chemical Problem
  • Fracture
  • Chemical Problem

Patient

NA · Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Additional Manufacturer Narrative

RESULT OF INVESTIGATION: THE DEVICE IN QUESTION (11272C1 CYSTO-URETHRO-FIBERSCOPE, SERIAL NUMBER: (B)(6), MANUFACTURED MARCH 14, 2024) WAS RECEIVED BY THE MANUFACTURING SITE IN GERMANY FOR INVESTIGATION ON OCTOBER 10, 2025. INVESTIGATION OF THE DEVICE REVEALED THAT THE SHAFT'S MARKING RINGS ARE DISCOLOURED AND THE SHEATHING OF THE SHAFT IS BRITTLE AND TORN IN SEVERAL PLACES. IN ADDITION, THE EYECUP HAS WHITE SPOTS. ACCORDING TO THE PRODUCT HISTORY REVIEW THE DEVICE WAS NOT REPAIRED OR MAINTAINED UNTIL TODAY. QUALITY TEST WAS PASSED AT THE TIME OF DELIVERY. IN THE PAST, A REPROCESSING AND INFORMATION CHECKLIST PROVIDED BY THE CUSTOMER WAS EVALUATED FOR SIMILAR COMPLAINTS WITH SIMILAR FAILURE DESCRIPTION. THE REASON FOR THE BURST OF THE SHAFT, IN EARLIER SIMILAR COMPLAINTS, WAS DUE TO NON-COMPLIANT REPROCESSING. ACCORDING TO THE INVESTIGATION THE FAILURE CAN BE TRACED BACK TO A REPROCESSING FAILURE. THE ABOVE CONDUCTED INVESTIGATIONS DID NOT REVEAL A ROOT CAUSE RELATED TO THE LABELLING, THE DEVICE OR ITS HISTORY. ALL PRODUCTION RELATED QUALITY CHECKS WERE PASSED, AND NO NON-CONFORMITIES WERE IDENTIFIED IN THE DEVICES MANUFACTURING RECORDS. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).

Additional Manufacturer Narrative

THE DEVICE IN QUESTION WAS RETURNED TO THE MANUFACTURER. THE EVALUATION IS ANTICIPATED, BUT NOT YET BEGUN. THE INVESTIGATION WILL BE PERFORMED BY A DESIGNATED KARL STORZ EMPLOYEE. THE EVENT IS FILED UNDER INTERNAL KARL STORZ COMPLAINT ID: (B)(4).

Description of Event or Problem

IT WAS REPORTED THE SHEATH OF THE DEVICE BROKE. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.