MalfunctionBRZ

CLEARLINK

Received Nov 14, 2025 · Event occurred Oct 17, 2025

Report 1416980-2025-05667 · MDR key 23552736

Device

Generic name

Set, Blood Transfusion

Model number

2C8750

Catalog number

2C8750

Lot number

DR25D07055

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: D4: EXPIRATION DATE (UPDATE THE N/A), D9, H3, H4, H6(UPDATE CODES), H11. H11: THE ACTUAL DEVICE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED AND OBSERVED THAT THE FILTER WAS BROKEN. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED, HOWEVER WAS MOST LIKELY RELATED TO PRODUCT HANDLING BY THE USER DURING THE INSTALLATION PROCESS. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT TWO (02) CLEARLINK SYSTEM Y-TYPE BLOOD/SOLUTION SETS WERE BROKEN AT THE CONNECTION POINT BETWEEN THE BLOOD FILTER CHAMBER AND THE TUBING. THIS WAS DISCOVERED DURING SETUP. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.