ROCHE 9180 ELECTROLYTE ANALYZER
Received Nov 10, 2025 · Event occurred Oct 15, 2025
Report 1823260-2025-04568 · MDR key 23512178
Device
Generic name
Electrolyte Analyzer
Manufacturer
Roche DiagnosticsCatalog number
03157334001
Lot number
2506106 192
Product problems
- Product Quality Problem
- Incorrect, Inadequate or Imprecise Result or Readings
- Product Quality Problem
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE ROOT CAUSE OF THE ISSUE WAS RELATED TO A REFERENCE ELECTRODE MANUFACTURED BY DIAMOND DIAGNOSTICS (DD REF ELECTRODE) USED WITH THE 9180 ELECTROLYTE ANALYZER. THE REFERENCE ELECTRODE USED (DD REF ELECTRODE) IS COVERED BY A ROCHE INITIATED RECALL. CUSTOMERS USING THE DD REF ELECTRODE WERE INSTRUCTED TO IMMEDIATELY STOP USING SODIUM (NA) RESULTS FROM THEIR 9180 ANALYZER. THE AFFECTED REFERENCE ELECTRODE WAS NOT DISTRIBUTED IN THE UNITED STATES. CUSTOMERS IN THE UNITED STATES USE ROCHE REFERENCE ELECTRODE AND REFERENCE ELECTRODE HOUSING. NO FURTHER ACTION IS NEEDED FOR CUSTOMERS IN THE UNITED STATES. MEDWATCH FIELDS D1, D2, D4, G1, AND G4 HAVE BEEN UPDATED.
Additional Manufacturer Narrative
THE SERIAL NUMBER OF THE 9180 INSTRUMENT IS (B)(6). THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH THE ROCHE 9180 SODIUM ELECTRODE ON A ROCHE 9180 ELECTROLYTE ANALYZER. THE SAMPLE WAS TESTED FIVE TIMES, RESULTING IN THE FOLLOWING SODIUM VALUES: 139 MMOL/L, 138 MMOL/L, 138 MMOL/L, 138 MMOL/L, AND 138 MMOL/L. THE SAMPLE WAS TESTED TWO ADDITIONAL TIMES ON A BACKUP 9180 INSTRUMENT, RESULTING IN SODIUM VALUES OF 114 MMOL/L WITH A DATA FLAG AND 112 MMOL/L WITH A DATA FLAG. THESE REPEAT VALUES WERE APPROPRIATE FOR THE PATIENT'S CLINICAL CONDITION.