inforMED
MalfunctionBWA

VALLEYLAB

Received Nov 10, 2025 · Event occurred Aug 18, 2025

Report 3026612481-2025-00022 · MDR key 23510567

Device

Generic name

Unit, Electrosurgical And Coagulation, With Accessories

Manufacturer

Covidien

Model number

E6008

Catalog number

E6008

Product problems

  • Energy Output Problem
  • Adverse Event Without Identified Device or Use Problem
  • No Apparent Adverse Event
  • Energy Output Problem
  • Adverse Event Without Identified Device or Use Problem
  • No Apparent Adverse Event

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT PRIOR TO USE, FLAME AND SMOKE WERE OBSERVED ON THE CORD CONNECTED TO THE MONOPOLAR 1 UNIVERSAL RECEPTACLE. THE CORD WAS BURNED AND BROKEN. THE UNIT'S CAUTERY SETTINGS WERE AS FOLLOWS: VALLEYLAB CAUTERY CUT 0 COAG 30. THE CORD USED WAS A NON-MEDTRONIC MANUFACTURED. THERE WAS NO PATIENT INVOLVED.

Additional Manufacturer Narrative

D10 CONCOMITANT PRODUCT: VLFT10GEN, VLFT10GEN FT SERIES ENERGY PLATFORMX1 (SERIAL#(B)(6)); E6009, E6009 BIP FOOTSW X1 (LOT#UNKNOWN). MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: G3, H6 ADDITIONAL INFORMATION RECEIVED SHOWS THAT THERE IS NO INFORMATION TO REASONABLY SUGGEST THAT THE DEVICE IN THIS REPORT MAY HAVE CAUSED OR CONTRIBUTED TO A DEATH OR SERIOUS INJURY OR THAT THE DEVICE IN THIS REPORT HAS MALFUNCTIONED. THEREFORE, THIS EVENT DOES NOT MEET THE REPORTING REQUIREMENTS IN 21 CFR 803. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.