ONE SNARE
Received Nov 7, 2025 · Event occurred Aug 8, 2025
Report MW5178685 · MDR key 23500485
Device
Generic name
Device, Percutaneous Retrieval
Manufacturer
Merit Medical Systems, Inc.Product problems
- Use of Device Problem
- Device Dislodged or Dislocated
Patient
NA · Unknown
- Low Blood Pressure/ Hypotension
- Cardiac Tamponade
- Cardiac Perforation
- Low Blood Pressure/ Hypotension
- Cardiac Tamponade
- Cardiac Perforation
Narrative
Description of Event or Problem
A LEAD EXTRACTION PROCEDURE COMMENCED TO REMOVE A RV LEAD DUE TO FRACTURE. THE LEAD WAS PREPPED WITH A SPECTRANETICS LLD TO PROVIDE TRACTION. BEGINNING WITH A SPECTRANETICS 13F TIGHTRAIL ROTATING DILATOR SHEATH, THE LEAD BROKE NEAR THE SVC COIL. THE TIGHTRAIL WAS REMOVED FROM THE BODY, AND THE DECISION WAS MADE TO SNARE FROM THE GROIN. A GORE MEDICAL 20F DRYSEAL FLEX INTRODUCER SHEATH WAS USED WITH AN ABBOTT AGILIS STEERABLE INTRODUCER. A MERIT MEDICAL ONE SNARE AND MERIT MEDICAL WHOLEY GUIDEWIRE WERE INITIALLY USED FOR SNARING. THE WHOLEY WIRE BROKE, SO THEN USED A BOSTON SCIENTIFIC SUPERSTIFF AMPLATZ GUIDEWIRE TO ATTEMPT TO GRAB THE LEAD. AFTER UNSUCCESSFUL ATTEMPTS TO PULL AND EXTRACT, SNARING CONTINUED WITH THE AMPLATZ WIRE AND MERIT MEDICAL ONE SNARE. THE LEAD WAS PULLED THROUGH THE SHEATH, THE ONE SNARE WAS DISENGAGED, AND WAS RAN OVER BOTH ENDS OF THE WIRE AT THE POINT OF THE LEAD. AFTER SUCCESSFULLY GRASPING THE LEAD, HE PULLED DOWN UNTIL THE LEAD DISLODGED FROM THE VASCULATURE. THE SVC COIL BROKE INTO TWO AND REMAINED IN THE SVC. THE REMAINING PORTION OF THE LEAD BROKE OFF AND STAYED INTACT AT THE DISTAL END. THEN, USING THE ONE SNARE, THE LOOSE END OF THE LEAD WAS PULLED UNTIL IT RELEASED. HE WAS ABLE TO REMOVE THE ENTIRE PORTION THROUGH THE SNARE AND DRYSEAL FLEW INTRODUCER SHEATH, OTHER THAN THE SVC COIL PORTION THAT HAD BROKEN OFF EARLIER IN THE CASE. ONCE RELEASED, PRESSURE DECLINED, AND AN RV PERFORATION WAS SUSPECTED. TEE WAS CHECKED AND IDENTIFIED PERICARDIA! TAMPONADE. PERICARDIOCENTESIS WAS PERFORMED AND ASPIRATED FLUID. THE BLOOD PRESSURE INCREASED AND THE INJURY BEGAN TO CLOT. ASPIRATION CONTINUED AND THE PATIENT STABILIZED. SHORTLY AFTER, THE BLOOD PRESSURE DECREASED AGAIN, AND CT SURGEON CAME IN TO LOOK AT THE CLOT AND DECIDED TO LEAVE AND MONITOR IN THE ICU. PATIENT WAS DISCHARGED TO THE ICU FOR MONITORING; THE NEEDLE WAS LEFT IN THE PATIENT IN CASE FURTHER ASPIRATION WAS NEEDED. PATIENT WAS STABLE, ALERT AND AWAKE WITHOUT THE NEED FOR PAIN MEDS A FEW HOURS FOLLOWING THE PERICARDIOCENTESIS. THE PATIENT SURVIVED THE PROCEDURE. THE PHYSICIAN STATED THAT THE ACT OF PULLING ON THE LEAD VIA FEMORAL SNARE ATTRIBUTED TO THE PERFORATION. THIS REPORT REFLECTS INFORMATION RECEIVED BY FDA IN THE FORM OF A NOTIFICATION PER 803.22 (B)(2).