inforMED
InjuryFCD

E-Z-EM BARIUM SOLUTION PREFILLED ENEMA BAG

Received Aug 5, 1999

Report MW1016891 · MDR key 234970

Device

Generic name

Enema Bag Clamp

Manufacturer

E-z-em, Inc.

Model number

NOT AVAILABLE

Catalog number

*

Lot number

*

Product problems

  • Loose

Patient

71 YR

  • Perforation

Narrative

Description of Event or Problem

DURING A WATER SOLUBLE BARIUM ENEMA, THERE WAS A PERFORATION OF THE PT'S COLON. DURING A RETROSPECTIVE REVIEW OF THIS INCIDENT, FINALIZED 5/24/99, IT WAS DETERMINED THAT THE ROOT CAUSE OF THE PERFORATION WAS A CLAMP ON THE BARIUM ENEMA BAG THAT DID NOT HOLD IN THE CLOSED POSITION. A LETTER WITH A COPY OF THE MEDWATCH FORM WILL BE SENT TO THE MFR. THE RADIOLOGY DEPT IS NOW USING TWO CLAMPS ON THE E-Z-EM ENEMA BAG, TO PREVENT ANY FUTURE OCCURRENCES.