PROVITA
Received Nov 6, 2025 · Event occurred Sep 30, 2025
Report 3002732397-2025-00002 · MDR key 23486864
Device
Generic name
Mobile Surgical Light
Manufacturer
Provita Medical Gmbh & Co. KgModel number
MRT LAMPCatalog number
350826
Lot number
23-2575
Product problems
- Fail-Safe Did Not Operate
- Fail-Safe Did Not Operate
Patient
NA · Unknown
- Dyspnea
- Dyspnea
Narrative
Description of Event or Problem
THIS IS A SURGICAL LAMP FOR USE IN MRI ENVIRONMENTS UP TO 3 TESLA. IT IS BATTERY OPERATED, EMPLOYING A LITHIUM ION RECHARGEABLE BATTERY. THE UNIT IS UL LISTED. WE HAVE BECOME AWARE OF A BATTERY FAILURE, DEVICE NAME: SERIES 1 MOBILE LED UL/MRI-LAMP WITH BATTERY. SYMPTOM: BATTERY OVERHEATED WHILE BEING CHARGED. THE OVERHEATING DAMAGED THE BATTERY HOUSING CAUSING CHARRING AND EMITTING SMOKE. ACCORDING TO THE USER FACILITY REPORTING THE INCIDENT, TWO PERSONS SUFFERED SMOKE INHALATION. BOTH PERSONS RETURNED TO WORK WITHOUT RECEIVING SERIOUS OR PERMANENT INJURY. NO PATIENTS WERE INVOLVED. WE HAVE BEEN UNABLE TO OBTAIN FULL DETAILS OF THE EVENT BECAUSE OF LACK OF DETAILED RESPONSES FROM THE USER FACILITY..
Additional Manufacturer Narrative
CORRECTIONS MADE: THIS IS SUPPLEMENT 1. CORRECTED RELATED REPORT NUIMBER.
Description of Event or Problem
THIS IS A SURGICAL LAMP FOR USE IN MRI ENVIRONMENTS UP TO 3 TESLA. IT IS BATTERY OPERATED, EMPLOYING A LITHIUM-ION RECHARGEABLE BATTERY. THE UNIT IS UL LISTED. WE HAVE BECOME AWARE OF A BATTERY FAILURE, DEVICE NAME: SERIES 1 MOBILE LED UL/MRI-LAMP WITH BATTERY. SYMPTOM: BATTERY OVERHEATED WHILE BEING CHARGED. THE OVERHEATING DAMAGED THE BATTERY HOUSING CAUSING CHARRING AND EMITTING SMOKE. ACCORDING TO THE USER FACILITY REPORTING THE INCIDENT, TWO PERSONS SUFFERED SMOKE INHALATION. SINCE TWO WERE AFFECTED, THERE ARE TWO MDR REPORTS. THIS IS REPORT NUMBER 1 OF 2. BOTH PERSONS RETURNED TO WORK WITHOUT RECEIVING SERIOUS OR PERMANENT INJURY. NO PATIENTS WERE INVOLVED. WE HAVE BEEN UNABLE TO OBTAIN FULL DETAILS OF THE EVENT BECAUSE OF LACK OF DETAILED RESPONSES FROM THE USER FACILITY.
Remedial action
- Replace