inforMED
MalfunctionLEG

ONLINE TDM VALPROIC ACID

Received Nov 3, 2025 · Event occurred Oct 8, 2025

Report 1823260-2025-04462 · MDR key 23453205

Device

Generic name

Enzyme Immunoassay, Valproic Acid

Manufacturer

Roche Diagnostics

Catalog number

04491041190

Lot number

815184

Product problems

  • Incorrect Measurement
  • Incorrect Measurement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION DETERMINED THE ISSUE WAS CONSISTENT WITH A PRE-ANALYTICAL HANDLING ISSUE.THE INVESTIGATION DID NOT IDENTIFY A SPECIFIC CAUSE OF THE EVENT.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ONLINE TDM VALPROIC ACID RESULTS FOR 1 PATIENT SAMPLE ON A COBAS 8000 COBAS C 502 MODULE. THE INITIAL RESULT WAS 4.2 G/ML. THE RESULT WAS QUESTIONED BY THE PHYSICIAN, AND THE SAMPLE WAS REPEATED. THE REPEATED RESULTS WERE 184.3 G/ML AND 109.5 G/ML. A DILUTION WAS PERFORMED, AND THE RESULT WAS 219 G/ML. THIS RESULT WAS BELIEVED TO BE CORRECT. THE CUSTOMER INDICATED THAT THE SAMPLE QUALITY WAS IN QUESTION.

Additional Manufacturer Narrative