inforMED
MalfunctionHNO

FS DISPOSABLE INTERFACE

Received Oct 31, 2025

Report 3012236936-2025-000283 · MDR key 23443056

Device

Generic name

Powered Laser Surgical Instrument

Model number

660065AN

Catalog number

660065AN

Lot number

60423182

Product problems

  • Decrease in Suction
  • Decrease in Suction

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEVICE EVALUATION: ONE (5) INVESTIGATION WAS COMPLETED DURING THE PERIOD. A VISUAL INSPECTION OF THE RETURNED PRODUCT DID NOT REVEAL ANY OBVIOUS DEFECTS OR DAMAGE. FUNCTIONALITY TEST WAS PERFORMED, AND THE RESULT WAS FOUND WITHIN SPECIFICATIONS. THE REPORTED EVENT IS CONFIRMED. A REVIEW OF THE RECORDS RELATED TO THE DEVICE THAT INCLUDED LABELING, MANUALS, TRENDING, AND DEVICE HISTORY RECORD (DHR) SHOWED THAT THE DEVICE AND ITS COMPONENTS MET ALL SPECIFICATIONS PRIOR TO BEING RELEASED. NO PRODUCT DEFICIENCY WAS IDENTIFIED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Additional Manufacturer Narrative

DEVICE EVALUATION: NO INVESTIGATIONS WERE COMPLETED DURING THE PERIOD. THE PATIENT INTERFACE (PI) SUCTION RING MAY LOSE SUCTION DURING A PROCEDURE. LABEL COPY STATES CORNEAL FIXATION VACUUM LOSS CAN OCCUR. THERE ARE SEVERAL FACTORS THAT MAY CONTRIBUTE TO SUCTION ISSUES SUCH AS DOCTOR¿S TECHNIQUE IN APPLYING THE SUCTION RING TO THE CORNEA, DOCTOR¿S TECHNIQUE IN SQUEEZING THE PI CLIP TO SECURE THE SUCTION RING TO THE PI CONE AND PATIENT ANATOMY AFFECTING THE INTERFACE BETWEEN THE PATIENT¿S CORNEA AND THE SUCTION RING. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.

Description of Event or Problem

THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS. THE EVENT WAS RELATED TO SUCTION LOSS WHILE LASER FIRING. THERE WERE NO PATIENT INJURIES REPORTED ASSOCIATED TO THE EVENTS.