UNICEL DXC 800 SYNCHRON SYSTEM

Received Nov 18, 2011 · Event occurred Oct 25, 2011

Report 2050012-2011-07622 · MDR key 2342607

Device

Generic name

Analyzer, Chemistry (photometric, Discrete), For Clinical Use

Model number

DXC 800 PRO

Catalog number

A11812

Product problems

  • Incorrect Or Inadequate Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE INSTRUMENT ISSUE WAS RESOLVED WHEN THE FIELD SERVICE ENGINEER REPLACED THE PHOSPHORUS CUP MODULE. (B)(4).

Description of Event or Problem

CUSTOMER CALLED TO REPORT THAT THE UNICEL DXC 800 SYNCHRON SYSTEM GENERATED FALSELY LOW PHOSPHORUS RESULTS FOR EIGHT PATIENTS. THE CUSTOMER TECHNICAL SPECIALIST REVIEWED CUSTOMER'S PROSERVICE ACCOUNT AND FOUND THAT THE QUALITY CONTROLS HAD ERRATIC RESULTS. ON THE FOLLOWING DAY, THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND REPLACED THE PHOSPHORUS CUP MODULE. THE FSE THEN VERIFIED INSTRUMENT PERFORMANCE TO ENSURE THAT THE SERVICES PERFORMED EFFECTIVELY RESOLVED THE INSTRUMENT ISSUE. CUSTOMER STATED THAT ALTHOUGH ERRONEOUS RESULTS WERE GENERATED, THEY WERE NOT REPORTED OUTSIDE THE LABORATORY; THEREFORE, PATIENT TREATMENT WAS NOT AFFECTED.