LEKSELL GAMMAPLAN
Received Oct 30, 2025 · Event occurred Sep 11, 2025
Report 3015232217-2025-00068 · MDR key 23422727
Device
Generic name
Radionuclide Rt Treatment Planning System
Manufacturer
Elekta Solutions AbLot number
11.4.1
Product problems
- Improper or Incorrect Procedure or Method
- Improper or Incorrect Procedure or Method
Patient
NA · Unknown
- Unintended Radiation Exposure
- Unintended Radiation Exposure
Narrative
Additional Manufacturer Narrative
THE MANUFACTURER'S INVESTIGATION IS ON-GOING AND FURTHER INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS COMPLETED.
Description of Event or Problem
THE CUSTOMER REPORTED A MISMATCH OF MR VS CBCT #1.
Additional Manufacturer Narrative
H11: UPDATED: THE INVESTIGATION WAS COMPLETED BY CONDUCTING A THOROUGH EVALUATION OF THE PRODUCT AND THE REPORTED INFORMATION. THE CUSTOMER REPORTED A MISMATCH OF MR VS CBCT #1. AN AVM TREATMENT WAS PLANNED USING A COMBINATION OF FRAME-BASED AND FRAMELESS IMAGES AND PLANNED FOR DELIVERY USING FOUR TREATMENTS. AFTER DELIVERY OF THE THIRD TREATMENT, IT WAS DISCOVERED THAT THE PLAN DID NOT CORRECTLY ALIGN ON THE STEREOTACTIC CBCT IMAGES. IT WAS CONCLUDED THAT THE PLAN RESULTED IN MISTREATMENT OF THE PATIENT WITH DOSE BEING DELIVERED IN THE WRONG POSITION. THE ROOT CAUSE IS THAT THE PLAN WAS NOT PROPERLY UPDATED AFTER CHANGING FIXATION. THE CLEAR DISCREPANCIES THAT THIS CAUSED WERE NOT PROPERLY REVIEWED OR ACTED UPON BEFORE THE APPROVAL OF THE PLAN. THE SYSTEM WORKED AS DESIGNED AND INTENDED. ELEKTA HAVE NOT BEEN PROVIDED WITH INFORMATION REGARDING THE PATIENT'S CONDITION. AN IMPORTANT FIELD SAFETY NOTICE (100-01-102-019) HAS BEEN SENT TO ALL AFFECTED CUSTOMERS FROM 17 DECEMBER 2025 ELEKTA REFERENCE #: (B)(4). TO REDUCE THE LIKELIHOOD FOR ERRONEOUS USE OF OBSOLETE STEREOTACTIC REFERENCE FOR TREATMENTS WITH A FRAME, LEKSELL GAMMAPLAN® 11.5 WILL NO LONGER PERMIT THE USE OF STEREOTACTIC CBCT REFERENCE IN COMBINATION WITH OTHER STEREOTACTIC REFERENCES.
Remedial action
- Modification/Adjustment
- Notification