VAPROX
Received Oct 29, 2025 · Event occurred Oct 14, 2025
Report 3003950207-2025-00010 · MDR key 23420499
Device
Generic name
Sterilant
Manufacturer
Steris CorporationModel number
PB011Catalog number
PB011
Lot number
(10)HC2305P1
Product problems
- Leak/Splash
- Leak/Splash
Patient
NA · Unknown
- Burn(s)
- Burn(s)
Narrative
Description of Event or Problem
THE USER FACILITY REPORTED AN EMPLOYEE OBTAINED A BURN ON THEIR HANDS WHILE HANDLING A LEAKING CARTON OF VAPROX HC STERILANT. NO MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED AND THE EMPLOYEE CONTINUED WORKING.
Additional Manufacturer Narrative
THE EMPLOYEE SUBJECT OF THE REPORTED EVENT WAS NOT WEARING PROPER PPE, SPECIFICALLY GLOVES, WHILE HANDLING THE VAPROC HC STERILANT. THE VAPROX HC STERILANT SAFETY DATA SHEET STATES (4), "HAND PROTECTION: WEAR PROTECTIVE GLOVES". THE SHIPPING CASE INSERT ALSO STATES, "WEAR PERSONAL PROTECTIVE EQUIPMENT TO HANDLE CARTRIDGE; CHEMICAL RESISTANT GLOVES". THE CAUSE OF THE REPORT LEAK COULD NOT BE DETERMINED; HOWEVER, IT IS LIKELY THAT THE PACKAGE WAS DAMAGED DURING SHIPPING AND TRANSPORTATION. THE DEVICE HISTORY RECORD WAS REVIEWED FOR THE SUBJECT LOT AND NO ABNORMALITIES WERE NOTED. A COMPLAINT REVIEW WAS PERFORMED AND CONFIRMED THE EVENT TO BE ISOLATED FOR THE SUBJECT LOT. USER FACILITY PERSONNEL WERE COUNSELED ON THE PROPER USE AND HANDLING FOR VAPROX HC STERILANT AND THE IMPORTANCE OF WEARING GLOVES. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.
Additional Manufacturer Narrative
INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION BECOMES AVAILABLE. IT SHOULD BE NOTED THAT THIS SPECIFIC MODEL IS NOT SOLD IN THE UNITED STATES, SO IT IS NOT IN GUDID.
Description of Event or Problem
THE USER FACILITY REPORTED AN EMPLOYEE OBTAINED A BURN ON THEIR HANDS WHILE HANDLING A LEAKING PACKAGE OF VAPROX HC STERILANT. IT IS UNKNOWN IF MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.