inforMED
InjuryLIO

APTIMA MULTITEST SPECIMEN COLLECTION KIT

Received Oct 28, 2025 · Event occurred Oct 13, 2025

Report 2024800-2025-00068 · MDR key 23407944

Device

Generic name

Device, Specimen Collection

Manufacturer

Hologic, Inc.

Catalog number

ASY-08848

Lot number

922935

Product problems

  • Material Separation
  • Material Separation

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

HOLOGIC REACHED OUT TO THE CUSTOMER MULTIPLE TIMES AND ATTEMPTED TO OBTAIN MORE INFORMATION ON THE SWAB LOT, COLLECTION METHOD, AND SAMPLE TYPE BUT DID NOT RECEIVE A RESPONSE. HOLOGIC EVALUATED RETENTION MATERIALS FROM THIS LOT LN 922935 BUT WAS NOT ABLE TO REPLICATE THE CUSTOMER ISSUE. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE OUTCOMES RELATED TO THIS ISSUE.

Description of Event or Problem

ON (B)(6) 2025, CUSTOMER REPORTED TO HOLOGIC THAT IN TWO SEPARATE PATIENT ENCOUNTERS, THE SWAB TIP FROM THE APTIMA MULTITEST SWAB SPECIMEN COLLECTION KIT (LN 922935) DETACHED FROM THE STICK AND REMAINED INSIDE THE PATIENT. CUSTOMER REPORTED THAT THE TWO PATIENT ENCOUNTERS OCCURRED WITHIN 24 HOURS PRIOR TO THEIR REPORT TO HOLOGIC. IN BOTH INSTANCES, THE PHYSICIAN WAS ABLE TO REMOVE THE SWAB TIP WITHOUT ANY REPORTED PATIENT HARM. CUSTOMER DID NOT PROVIDE PATIENT INFORMATION OR CONFIRMATION OF SWAB SPECIMEN TYPE OR COLLECTION METHOD. THIS REPORT IS TO DOCUMENT THE FIRST AFFECTED PATIENT ENCOUNTER. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE OUTCOMES RELATED TO THIS EVENT.