4.5MM VA-LCP CURVED CONDYLAR PLATE/10HOLE/230MM/RT-STER
Received Oct 28, 2025 · Event occurred May 30, 2025
Report 8030965-2025-10776 · MDR key 23404579
Device
Generic name
Implant, Fixation Device, Condylar Plate
Manufacturer
Synthes GmbhCatalog number
02.124.410S
Lot number
30289P9
Product problems
- Break
- Break
Patient
NA · Unknown
- Pain
- Pain
Narrative
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11. ADDITIONAL NARRATIVE: ADDED: D9. H3, H6: THE DEVICE WAS RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
Additional Manufacturer Narrative
PRODUCT COMPLAINT #: (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.
Description of Event or Problem
IT WAS REPORTED THAT THE PATIENT WAS OPERATED ON FOR A SUPRA- AND INTER-CONDYLAR FRACTURE OF THE RIGHT FEMUR (OSTEOSYNTHESIS), WHO CONSULTS URGENTLY FOR PAIN IN THE FEMORAL SHAFT. THE XRAY REVEALS A FRACTURE OF THE CONDYLAR PLATE OF THE RIGHT FEMUR, IMPLANTED DURING THE INTERVENTION.
Additional Manufacturer Narrative
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: H3, H6: THE PRODUCT WAS RETURNED TO J&J MEDTECH ORTHOPAEDICS FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THAT THE VA-LCP CONDYLAR PLATE 4.5/5.0 R 10HO L23 WAS BROKEN FROM ONE OF THE HOLES. THE BROKEN FRAGMENT WAS RETURNED FOR EXAMINATION AND THE FRACTURE SURFACE WAS EXAMINED, AND THE FRACTURE SUGGEST THAT THE IMPLANT WAS EXPOSED TO A BIG STRESS CAUSING THE RUPTURE OF THE IMPLANT. OTHER SECTION OF THE FRACTURE SHOW SMALL AREAS IN THE FORM OF BEACH MARKS, INDICATING FATIGUE. THEREFORE, BASED ON THIS EVIDENCE IT IS REASONABLE TO CONCLUDE THAT THE FRACTURE WAS CAUSED IN TWO EVENTS THE FIRST WERE THE IMPLANT UNDERWENT A HIGH STRESS AN THE SECOND THE LOW CYCLE FATIGUE. THERE IS NO INDICATION OF DAMAGE RELATED TO MATERIAL ISSUES WAS OBSERVED. ADDITIONALLY IT WAS OBSERVED WITH SIGNS OF SCRATCHES AND WEAR WHICH COULD HAVE BEEN DUE TO IMPLANTATION AND EXTRACTION PROCESS. WITH THE INFORMATION PROVIDED IS NOT POSSIBLE TO DETERMINE A POTENTIAL CAUSE AT THIS MOMENT. THE MODE OF FAILURE OF THE DEVICE IS MULTI-FACTORIAL AND CONSIDERATION MUST BE GIVEN TO ALL OTHER POTENTIAL INFLUENCES SUCH AS SURGICAL PROCESS, PATIENT VARIABLES (E.G. ACTIVITY LEVEL AND USE), ANATOMICAL CONSIDERATIONS AND PATIENT CHANGES OVER TIME. ANY CONCLUSIONS FROM THE INVESTIGATIONAL INPUT PROVIDED HAVE TO BE PLACED INTO CONTEXT WITH ALL OTHER RELEVANT FACTORS. A DIMENSIONAL INSPECTION WAS UNABLE TO BE PERFORMED DUE TO THE POST-MANUFACTURING DAMAGE. A FUNCTIONAL TEST WAS UNABLE TO BE PERFORMED DUE TO THE POST-MANUFACTURING DAMAGE THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE VA-LCP CONDYLAR PLATE 4.5/5.0 R 10HO L23 WOULD HAVE CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT ==> PART NUMBER: 02.124.410S LOT NUMBER: 30289P9 MANUFACTURING SITE: (B)(4). RELEASE TO WAREHOUSE DATE: 04 SET 2024 EXPIRATION DATE: 01 AUG 2034 A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE STERILE FINISHED LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.