inforMED
MalfunctionJCF

BD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML

Received Oct 23, 2025 · Event occurred Sep 26, 2025

Report 1917413-2025-01463 · MDR key 23369687

Device

Generic name

Lymphocyte Separation Medium

Catalog number

362780

Lot number

4346567

Product problems

  • Fluid/Blood Leak
  • Fluid/Blood Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED ANY SAMPLES OR PHOTOS FOR INVESTIGATION. THE RETAINED SAMPLES FROM LOT 4346567, TOTALING 50, WERE VISUALLY INSPECTED WITH NO ISSUES IDENTIFIED. SIMILARLY, THE RETAINED SAMPLES FROM LOT 5107761, TOTALING 60, WERE ALSO VISUALLY INSPECTED WITH NO ISSUES IDENTIFIED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOTS, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE LOTS 4346567 AND 5107761, FOR THE INDICATED FAILURE MODE: GLASS BREAKAGE DURING CENTRIFUGATION. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem

REPORT 1 OF 2: IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML, AN UNSPECIFIED NUMBER OF TUBES BROKE DURING CENTRIFUGATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.