NEW IMAGE DRAINABLE POUCH
Received Oct 23, 2025 · Event occurred Oct 3, 2025
Report 1119193-2025-00041 · MDR key 23366856
Device
Generic name
2-piece Drainable Ostomy Pouch
Manufacturer
Hollister IncorporatedModel number
18003Catalog number
18003
Lot number
5F252
Product problems
- Difficult to Open or Close
- Difficult to Open or Close
Patient
74 YR · Female
- Urinary Tract Infection
- Urinary Tract Infection
Narrative
Additional Manufacturer Narrative
THE LOT NUMBER WAS PROVIDED, ENABLING A DEVICE HISTORY RECORD (DHR) REVIEW, WHICH WAS COMPLETED WITH NO ANOMALIES IDENTIFIED. HOWEVER, THE POUCH WAS NOT RETURNED; THEREFORE, A DEVICE EVALUATION COULD NOT BE CONDUCTED. A COMPLAINT HISTORY TREND ANALYSIS WAS PERFORMED FOR SIMILAR COMPLAINTS, AND NO ADVERSE TRENDS WERE OBSERVED. BASED ON THE INFORMATION AVAILABLE, A ROOT CAUSE COULD NOT BE DETERMINED. HOLLISTER WILL CONTINUE TO MONITOR THIS REPORTED ISSUE.
Description of Event or Problem
IT WAS REPORTED THAT AN END USER EXPERIENCED EFFLUENT LEAKAGE WHEN THE LOCK N ROLL CLOSURE OF THE HOLLISTER NEW IMAGE DRAINABLE POUCH UNROLLED. IT WAS REPORTED THAT THE END USER WEARS DEPENDS, WHICH ALSO COVERS THE POUCH, AND THE POUCH HAS BEEN OPENING, RESULTING IN CONTENTS LEAKING OUT INTO THE DEPENDS. IT WAS FURTHER REPORTED THAT THE LEAKAGE INTO HER DEPENDS ON LED TO A URINARY TRACT INFECTION (UTI), FOR WHICH SHE REQUIRED ANTIBIOTIC TREATMENT.