IPS PRESSURE SENSOR CONNECTING CABLE
Received Oct 23, 2025 · Event occurred Oct 8, 2025
Report 3012172416-2025-00102 · MDR key 23366760
Device
Generic name
Extracorporeal System For Long-term Respiratory / Cardiopulmonary Failure
Manufacturer
Xenios AgCatalog number
F38350001
Lot number
8013116
Product problems
- Display or Visual Feedback Problem
- Incorrect, Inadequate or Imprecise Result or Readings
- Output Problem
- Display or Visual Feedback Problem
- Incorrect, Inadequate or Imprecise Result or Readings
- Output Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: B5, D4, D9, H3, H6 PLANT INVESTIGATION: NONCONFORMITY WAS NOT OBSERVED DURING THE MANUFACTURING PROCESS. ONE SIMILAR COMPLAINT HAS BEEN REPORTED CONCERNING THIS SERIAL NUMBER. THE PRODUCT WAS NOT RETURNED, AND A PHYSICAL INVESTIGATION COULD NOT BE CARRIED OUT. A DEFECT IN THE CABLE OR CABLE CONNECTION MAY HAVE CAUSED THE ISSUE, HOWEVER, A DEFINITIVE CAUSE FOR THE DEFECT COULD NOT BE DETERMINED. THE PROBLEM DESCRIBED IS POTENTIALLY ATTRIBUTED TO A KNOWN ISSUE. THE PLUG CONNECTION TO THE SENSOR MUST HAVE A TIGHT FIT TO ENSURE THAT IT IS WATERTIGHT. HOWEVER, THE DESIGN OF THE CABLE AND THE RESULTING MANUFACTURING PROCESS MAKES THE CABLE SUSCEPTIBLE TO DAMAGE DURING USE (BREAKING OR TEARING OUT OF THE COMPONENTS). THE COMPONENT IS A PURCHASED PART. THE ISSUE HAS BEEN FORWARDED TO RESEARCH AND DEVELOPMENT TO REVIEW THE DESIGN FOR IMPROVEMENT.
Description of Event or Problem
A USER FACILITY REPORTED WHEN A PATIENT WAS PLACED ON EXTRACORPOREAL MEMBRANE OXYGENATION SUPPORT UTILIZING PUMP #1, THE USER FOUND ONE OF THE PRESSURE CABLES WAS NOT READING. THE USER HAD A REPLACEMENT PRESSURE CABLE AND THEY WERE ABLE TO SWAP IT OUT WHICH RESOLVED THE ISSUE. THERE WAS NO INDICATION OF PATIENT SERIOUS INJURY. THE COMPLAINT SAMPLE WAS AVAILABLE FOR PRODUCT EVALUATION; HOWEVER, THE COMPLAINT SAMPLE WAS NOT RECEIVED.
Additional Manufacturer Narrative
THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.
Description of Event or Problem
A USER FACILITY REPORTED WHEN A PATIENT WAS PLACED ON EXTRACORPOREAL MEMBRANE OXYGENATION SUPPORT UTILIZING PUMP #1, THE USER FOUND ONE OF THE PRESSURE CABLES WAS NOT READING. THE USER HAD A REPLACEMENT PRESSURE CABLE AND THEY WERE ABLE TO SWAP IT OUT WHICH RESOLVED THE ISSUE. THERE WAS NO INDICATION OF PATIENT SERIOUS INJURY. THE COMPLAINT SAMPLE WAS AVAILABLE FOR PRODUCT EVALUATION.