CARDIOCHEK PLUS ANALYZER
Received Oct 22, 2025 · Event occurred Sep 22, 2025
Report 1836135-2025-00017 · MDR key 23363831
Device
Generic name
Clinical Chemistry Analyzer
Manufacturer
Pts DiagnosticsProduct problems
- Electrical /Electronic Property Problem
- Electrical /Electronic Property Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
CURRENTLY IT IS UNKNOWN WHETHER THE DEVICE MAY HAVE MALFUNCTIONED, AS THE ALLEGATION COULD NOT BE DUPLICATED. IT IS KNOWN THAT IMPROPER INSERTION OF BATTERIES - IN ANY DEVICE - MAY RESULT IN OVERHEATING, BATTERY LEAKAGE, ETC., WHICH IS STATED IN THE LABELING OF MOST BATTERY MANUFACTURERS. THE REPORTED DEVICE WAS BEING USED BEYOND IT'S 3-YEAR SHELF LIFE, AND SUBSEQUENT EXAMINATION OF THE RETURNED ANALYSER LED TO OBSERVATIONS CONSISTENT WITH DEVICE MISHANDLING.
Description of Event or Problem
THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THE CUSTOMER REPORTED THAT AFTER BATTERY INSTALLATION, THEIR DEVICE WAS"HOT TO THE TOUCH AND THE BATTERIES APPEARED TO BE MELTING.". THE REPORTED ANALYZER WAS BEING USED BEYOND IT'S STATED 3-YEAR SHELF LIFE.