inforMED
MalfunctionFPB

NA

Received Oct 21, 2025 · Event occurred Sep 21, 2025

Report 1416980-2025-05196 · MDR key 23348095

Device

Generic name

Filter, Infusion Line

Model number

2N3350

Catalog number

2N3350

Lot number

DR24K15031

Product problems

  • Crack
  • Fluid/Blood Leak
  • Crack
  • Fluid/Blood Leak

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem

IT WAS REPORTED THAT A MICRO-VOLUME EXTENSION SET LEAKED FROM A CRACK IN THE FILTER. THIS WAS OBSERVED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: H11: THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. PRESSURE TEST WAS PERFORMED, AND NO LEAK WAS OBSERVED. THE REPORTED CONDITION WAS NOT VERIFIED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.