Received Oct 21, 2025 · Event occurred Sep 21, 2025
Report 1416980-2025-05196 · MDR key 23348095
Device
Generic name
Filter, Infusion Line
Manufacturer
Baxter Healthcare CorporationModel number
2N3350Catalog number
2N3350
Lot number
DR24K15031
Product problems
- Crack
- Fluid/Blood Leak
- Crack
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THAT A MICRO-VOLUME EXTENSION SET LEAKED FROM A CRACK IN THE FILTER. THIS WAS OBSERVED DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION: H11: THE DEVICE WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. PRESSURE TEST WAS PERFORMED, AND NO LEAK WAS OBSERVED. THE REPORTED CONDITION WAS NOT VERIFIED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.