RESUSCITATOR,ADLT,MASK,TUB RES
Received Oct 20, 2025 · Event occurred Sep 15, 2025
Report 1417592-2025-00561 · MDR key 23336402
Device
Generic name
Resuscitator,adlt,mask,tub Res
Manufacturer
Medline Industries, LpCatalog number
CPRM4426
Lot number
561911
Product problems
- Decrease in Pressure
- Decrease in Pressure
Patient
NA · Male
- Low Oxygen Saturation
- Low Oxygen Saturation
Narrative
Description of Event or Problem
DURING A RESUSCITATION ATTEMPT INVOLVING AN ADULT MALE PATIENT, THE MEDLINE RESUSCITATION BAG REPORTEDLY FAILED TO ADEQUATELY VENTILATE THE PATIENT. THE PATIENT EXPERIENCED DESATURATION AND DID NOT IMPROVE WHILE THE MEDLINE DEVICE WAS IN USE. UPON SWITCHING TO A PREVIOUSLY STOCKED RESUSCITATION BAG FROM ANOTHER BRAND, VENTILATION IMPROVED AND THE PATIENT'S OXYGEN SATURATION INCREASED. THE FACILITY REPORTED PROBLEMS WITH ADEQUATE DELIVERY OF POSITIVE PRESSURE WHEN THE BAG IS NOT HELD PERFECTLY STRAIGHT, AS PRESSURE IS LOST WHEN TORQUE IS APPLIED AT THE JUNCTION TO THE ADAPTER.
Additional Manufacturer Narrative
DURING A RESUSCITATION ATTEMPT INVOLVING AN ADULT MALE PATIENT, THE MEDLINE RESUSCITATION BAG REPORTEDLY FAILED TO ADEQUATELY VENTILATE THE PATIENT. THE PATIENT EXPERIENCED DESATURATION AND DID NOT IMPROVE WHILE THE MEDLINE DEVICE WAS IN USE. UPON SWITCHING TO A PREVIOUSLY STOCKED RESUSCITATION BAG FROM ANOTHER BRAND, VENTILATION IMPROVED AND THE PATIENT'S OXYGEN SATURATION INCREASED. THE FACILITY REPORTED PROBLEMS WITH ADEQUATE DELIVERY OF POSITIVE PRESSURE WHEN THE BAG IS NOT HELD PERFECTLY STRAIGHT, AS PRESSURE IS LOST WHEN TORQUE IS APPLIED AT THE JUNCTION TO THE ADAPTER. IT HAS BEEN DETERMINED THAT THE REPORTED EVENT CAUSED OR CONTRIBUTED TO SERIOUS INJURY REQUIRING MEDICAL INTERVENTION, THEREFORE, THIS MEDWATCH IS BEING FILED. IF ANY FURTHER RELEVANT INFORMATION IS IDENTIFIED OR OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED.