APTIMA COMBO 2 ASSAY
Received Oct 17, 2025 · Event occurred Sep 24, 2025
Report 2024800-2025-00065 · MDR key 23324701
Device
Generic name
Dna-reagents, Neisseria
Manufacturer
Hologic, Inc.Model number
COMMONCatalog number
ASY-13955
Lot number
915869, 922619
Product problems
- Off-Label Use
- Off-Label Use
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
HOLOGIC TECHNICAL SUPPORT (TS) REVIEWED THE LOGS AND WORKLISTS AND FOUND NO EVIDENCE OF RELEVANT HARDWARE ERRORS OR THAT THERE WERE REAGENT PREPARATION OR CONTAMINATION ISSUES. TS ADVISED CUSTOMER THAT PER THE AC2 PACKAGE INSERT, THE FIRST VALID RESULT SHOULD BE REPORTED. HOLOGIC PERFORMED A RISK ASSESSMENT AND NOTED NO PRODUCT IMPACT. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE PATIENT OUTCOMES RELATED TO THIS ISSUE.
Description of Event or Problem
ON (B)(6) 2025, A CUSTOMER REPORTED A DISCREPANT RESULT USING THE APTIMA COMBO 2 (AC2) ASSAY ON THEIR PANTHER INSTRUMENT. ON (B)(6) 2025, SAMPLE ID (B)(6) RESULTED AS CT POSITIVE, GC NEGATIVE WITH MASTER LOT 915869 ON WORKLIST (B)(4). CUSTOMER REPORTED THEY HAVE MADE IT THEIR LAB POLICY TO RETEST ANY POSITIVE RESULT WITH AN RLU VALUE BELOW 1000. ON (B)(6) 2025, CUSTOMER RERAN THE SAMPLE USING MASTER LOT 922619 ON WORKLIST (B)(4) WHERE IT RESULTED AS DOUBLE CT NEGATIVE / GC NEGATIVE. CUSTOMER RELAYED THE INITIAL POSITIVE RESULT WAS HELD BACK, BUT DUE TO HOW THEIR LIS WAS PROGRAMMED, THE NEGATIVE RESULT WAS TRANSMITTED TO THE CAREGIVER. HOLOGIC HAS NOT LEARNED OF ANY ADVERSE PATIENT OUTCOMES DUE TO THIS EVENT.