inforMED
MalfunctionDXO

COMET II

Received Oct 16, 2025 · Event occurred Jul 14, 2025

Report 2124215-2025-71978 · MDR key 23310865

Device

Generic name

Transducer, Pressure, Catheter Tip

Model number

2404-02

Catalog number

2404-02

Lot number

0036026328

Product problems

  • Communication or Transmission Problem
  • Adverse Event Without Identified Device or Use Problem
  • Communication or Transmission Problem
  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

DEVICE EVALUATED BY MANUFACTURER. THE COMPLAINT DEVICE WAS RECEIVED FOR PRODUCT ANALYSIS. INSPECTION OF THE DEVICE REVEALED THAT THERE WERE NUMEROUS BENDS AND KINKS THROUGHOUT THE BODY FROM DISTAL TO PROXIMAL, ADDITIONALLY THE WIRE RETURNED DETACHED FROM THE BODY AT 39 CM FROM DISTAL TO PROXIMAL. NO OTHER ISSUES WERE IDENTIFIED DURING THE PRODUCT ANALYSIS.

Description of Event or Problem

REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 19SEP2025. IT WAS REPORTED THAT THERE WAS NO IMAGE. A COMET II PRESSURE GUIDEWIRE WAS SELECTED FOR USE. IT WAS REPORTED THAT THERE WAS NO IMAGE. HOWEVER, DEVICE ANALYSIS REVEALED THAT A PINHOLE WAS NOTED ON THE OUTER AT 5.2CM FROM THE TIP.