COMET II

Received Oct 15, 2025 · Event occurred Sep 23, 2025

Report 2124215-2025-72496 · MDR key 23304702

Device

Generic name

Transducer, Pressure, Catheter Tip

Model number

2404-02

Catalog number

2404-02

Lot number

0035907115

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

78 YR · Female

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

GOOD FAITH EFFORT ATTEMPTS WERE MADE TO TRY AND RETRIEVE ADDITIONAL DETAILS REGARDING THE REPORTED EVENT, BUT FURTHER INFORMATION WAS UNABLE TO BE OBTAINED. INVESTIGATION RESULTS: DEVICE ANALYSIS FINDINGS: THE DEVICE WAS NOT RETURNED FOR ANALYSIS; THEREFORE, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. DEVICE HISTORY RECORD (DHR) REVIEW: IT WAS CONFIRMED THAT THIS DEVICE MET MANUFACTURING SPECIFICATION PRIOR TO DISTRIBUTION AND THERE NO MANUFACTURING DEVIATIONS WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. INVESTIGATION CONCLUSION: THIS INVESTIGATION HAS BEEN ASSIGNED THE CONCLUSION CODE OF ADVERSE EVENT RELATED TO PROCEDURE FOLLOWING A REVIEW OF THE REPORTED EVENT DETAILS, AVAILABLE INFORMATION, AND PRODUCT RECORD REVIEW. IT IS LIKELY THAT THE ISSUE COULD BE RELATED TO OPERATIONAL FACTORS SUCH AS HANDLING, TECHNIQUE, AMONG OTHERS THAT CAUSED THE REPORTED EVENT, FURTHERMORE, BATCH RECORD REVIEW CONCLUDED THAT NO MANUFACTURING DEVIATIONS WERE IDENTIFIED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem

IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. A COMET II GUIDEWIRE WAS SELECTED FOR USE. DURING THE PROCEDURE, THE TIP OF THE DEVICE WAS BROKEN UPON INSERTION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.

Additional Manufacturer Narrative

GOOD FAITH EFFORT ATTEMPTS WERE MADE TO TRY AND RETRIEVE ADDITIONAL DETAILS REGARDING THE REPORTED EVENT, BUT FURTHER INFORMATION WAS UNABLE TO BE OBTAINED.

Description of Event or Problem

IT WAS REPORTED THAT TIP DETACHMENT OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ANTERIOR DESCENDING ARTERY. A COMET II GUIDEWIRE WAS SELECTED FOR USE. DURING THE PROCEDURE, THE TIP OF THE DEVICE WAS BROKEN. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.