HYDROS ROBOTIC SYSTEM
Received Oct 15, 2025 · Event occurred Sep 30, 2025
Report 3012977056-2025-00261 · MDR key 23301973
Device
Generic name
Fluid Jet Removal System
Manufacturer
Procept Biorobotics CorporationCatalog number
HY1000T
Product problems
- Adverse Event Without Identified Device or Use Problem
- Adverse Event Without Identified Device or Use Problem
Patient
NA · Male
- Hemorrhage/Blood Loss/Bleeding
- Hemorrhage/Blood Loss/Bleeding
Narrative
Additional Manufacturer Narrative
THE HYDROS ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS CURRENTLY IN THE POSSESSION OF THE USER FACILITY. THE INVESTIGATION OF THIS EVENT CONSISTED OF A REVIEW OF THE TREATMENT LOG FILES, DEVICE HISTORY RECORD (DHR) AND USER MANUAL (UM). THE HYDROS ROBOTIC SYSTEM'S TREATMENT LOG FILES WERE REVIEWED, WHICH CONFIRMED NO MALFUNCTION OCCURRED. THE REVIEW OF THE TREATMENT LOG FILES REVEALED THAT THE HYDROS ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AS NO MALFUNCTION WAS OBSERVED DURING AQUABLATION THERAPY. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR HY1000T/SERIAL NUMBER (B)(6) WAS CONDUCTED, WHICH CONFIRMED THAT THERE WERE NO NON-CONFORMANCES, FAILURES, DISCREPANCIES, OR MISSED STEPS DURING THE MANUFACTURING PROCESS THAT COULD BE RELATED TO THE REPORTED EVENT. THE REVIEW INDICATED THAT THE SYSTEM MET ALL DESIGN AND MANUFACTURING SPECIFICATIONS WHEN RELEASED FOR DISTRIBUTION. THE HYDROS ROBOTIC SYSTEM USER MANUAL, US, ENGLISH, UM0401-00-01 REV. C, WAS REVIEWED. 4.2.3 WARNINGS: PROCEDURE: AS WITH ANY SURGICAL UROLOGIC PROCEDURE, POTENTIAL PERIOPERATIVE RISKS OF THE AQUABLATION PROCEDURE INCLUDE THE FOLLOWING, SOME OF WHICH MAY LEAD TO SERIOUS OUTCOMES AND MAY REQUIRE INTERVENTION: BLEEDING OR BLOOD IN THE URINE. SECTION 9: POST-OP STERILE, HEMOSTASIS STATES THE FOLLOWING: PRIOR TO INTRODUCING THE RESECTOSCOPE TO START HEMOSTASIS, REDUCE TRUS COMPRESSION AND ENSURE THERE IS NO STEEP ANGLING OF THE TRUS PROBE. NOTE: DEPENDING ON PATIENT ANATOMY, CONSIDER THAT TRUS PROBE MAY BE REMOVED ENTIRELY, IF NEEDED TO MANIPULATE THE RESECTOSCOPE ADEQUATELY FOR HEMOSTASIS. WARNING: AVOID APPLYING EXCESSIVE COMPRESSION TO THE PROSTATE. EXCESSIVE COMPRESSION CAN CONTRIBUTE TO SERIOUS INJURY TO THE PATIENT. PREPARE THE RESECTOSCOPE AND TOOMEY SYRINGE/CLOT EVACUATOR FOR HEMOSTASIS. INTRODUCE RESECTOSCOPE SHEATH. EVACUATE CLOTS WITH TOOMEY OR ELLIK EVACUATOR. CAUTION: POSITION THE ELECTROSURGICAL GENERATOR APART FROM THE HYDROS ROBOTIC SYSTEM, AS PERMITTED BY THE OPERATING ROOM LAYOUT. USE A LOOP TO SYSTEMATICALLY WORK AROUND THE BLADDER NECK AND REMOVE THE ¿FLUFFY¿ TISSUE (TISSUE DEVITALIZED BY AQUABLATION THERAPY) TO VIEW POTENTIAL BLEEDING VESSELS UNDERNEATH. PERFORM FOCAL CAUTERY AT BLEEDERS. TURN OFF IRRIGATION AND INSPECT FOR BLEEDERS UNDER NORMAL BLOOD PRESSURE. NOTE: ALTERNATE HEMOSTASIS METHODS MAY INCLUDE: - BALLOON CATHETER IN BLADDER WITH BLADDER NECK TRACTION. - BALLOON CATHETER IN BLADDER WITH BLADDER NECK TRACTION. FILL THE BLADDER WITH STERILE SALINE AND MAINTAIN FOR 30-60 MINUTES BEFORE STARTING CBI. UNDER SPINAL ANESTHESIA. - BALLOON CATHETER IN BLADDER, NO TRACTION. - BALLOON CATHETER IN PROSTATIC FOSSA. - INFLATE BALLOON WITH 5CC SALINE IN THE BLADDER. - RETRACT BALLOON UNDER TRUS GUIDANCE INTO PROSTATIC FOSSA. - INFLATE BALLOON TO 30-50% OF THE INITIAL PROSTATE VOLUME. - APPLY MILD TRACTION TO HOLD THE BALLOON CATHETER IN PLACE. ONCE HEMOSTASIS IS ACHIEVED, REMOVE TRUS PROBE AND INSPECT PROBE AND ANUS FOR BLOOD AND INVESTIGATE, AS NEEDED. INTRODUCE 22-24F HEMATURIA CATHETER, INFLATE BALLOON, AND APPLY STANDARD TO LIGHT TRACTION AS NEEDED. RUN CBI PER FACILITY'S TURP PROTOCOL. MONITOR IRRIGATION BAGS TO MAINTAIN FLUID SUPPLY. MONITOR EFFLUENT COLOR. WARNING: UNADDRESSED VASCULAR BLEEDING CAN LEAD TO SERIOUS PATIENT HARM. THE HYDROS ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS STILL CURRENTLY IN POSSESSION OF THE USER FACILITY. IT WAS REPORTED THAT THE SURGEON HAD DIFFICULTY ACHIEVING HEMOSTASIS AND SUBSEQUENTLY A BLOOD TRANSFUSION WAS ADMINISTERED DUE TO A DROP IN HEMOGLOBIN. THE PATIENT IS DOING WELL. NO DEVICE MALFUNCTION WAS REPORTED. THE HYDROS ROBOTIC SYSTEM'S USER MANUAL (UM) LISTS BLEEDING AS A POTENTIAL RISK OF AQUABLATION THERAPY AND PROVIDES ADEQUATE INSTRUCTIONS ON HOW TO ACHIEVE APPROPRIATE HEMOSTASIS. BASED ON THE REVIEW OF THE INFORMATION PROVIDED, PLUS A REVIEW OF THE TREATMENT LOG FILES, DHR, AND UM, THE EVENT IS CONSIDERED NOT TO BE DEVICE RELATED. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
Description of Event or Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BECAME AWARE THAT POST-AQUABLATION THERAPY, THE TREATING SURGEON EXPERIENCED DIFFICULTY ACHIEVING HEMOSTASIS, AND SUBSEQUENTLY, A BLOOD TRANSFUSION WAS ADMINISTERED DUE TO A DROP IN HEMOGLOBIN LEVELS. THE PATIENT IS DOING WELL. NO MALFUNCTION OF THE HYDROS ROBOTIC SYSTEM WAS REPORTED.