DECOMPRESSION OF CHOICE
Received Oct 15, 2025 · Event occurred Sep 30, 2025
Report 3007278668-2025-00002 · MDR key 23291041
Device
Product problems
- Mechanical Problem
- Mechanical Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
A WELD ON PART E7130 APPEARS TO HAVE FAILED AND THE PIECE OF THE TABLE CONNECTED THROUGH THE E7130 WELD FELL TO THE GROUND.
Additional Manufacturer Narrative
THE DEVICE IS BEING RETURNED TO MANUFACTURER FOR EVALUATION. EVALUATION OF INITIAL INFORMATION APPEARS IT WAS LIKELY THERE WAS A WELD FAILURE.
Description of Event or Problem
A WELD ON PART E7130 APPEARS TO HAVE FAILED AND THE PIECE OF THE TABLE CONNECTED THROUGH THE E7130 WELD FELL TO THE GROUND.
Additional Manufacturer Narrative
THE DEVICE AND PART E7130 WERE EVALUATED AND A DEFECT IN THE WELD WAS IDENTIFIED AS THE ROOT CAUSE FOR THE MALFUNCTION THAT OCCURRED IN THIS REPORTED INCIDENT.
Remedial action
- Replace