inforMED
MalfunctionITH

DECOMPRESSION OF CHOICE

Received Oct 15, 2025 · Event occurred Sep 30, 2025

Report 3007278668-2025-00002 · MDR key 23291041

Device

Generic name

Doc

Model number

E9011

Product problems

  • Mechanical Problem
  • Mechanical Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

A WELD ON PART E7130 APPEARS TO HAVE FAILED AND THE PIECE OF THE TABLE CONNECTED THROUGH THE E7130 WELD FELL TO THE GROUND.

Additional Manufacturer Narrative

THE DEVICE IS BEING RETURNED TO MANUFACTURER FOR EVALUATION. EVALUATION OF INITIAL INFORMATION APPEARS IT WAS LIKELY THERE WAS A WELD FAILURE.

Description of Event or Problem

A WELD ON PART E7130 APPEARS TO HAVE FAILED AND THE PIECE OF THE TABLE CONNECTED THROUGH THE E7130 WELD FELL TO THE GROUND.

Additional Manufacturer Narrative

THE DEVICE AND PART E7130 WERE EVALUATED AND A DEFECT IN THE WELD WAS IDENTIFIED AS THE ROOT CAUSE FOR THE MALFUNCTION THAT OCCURRED IN THIS REPORTED INCIDENT.

Remedial action

  • Replace