STLESS STEEL SIL 4X18IN 1 S/A V-37
Received Oct 15, 2025 · Event occurred Jan 1, 2025
Report 2210968-2025-11747 · MDR key 23290034
Device
Product problems
- Material Separation
- Material Separation
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE NEEDLES ARE DETACHING FROM THE SUTURE. THERE WERE NO PATIENT ADVERSE EVENT. PRODUCT IS AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Additional Manufacturer Narrative
PRODUCT COMPLAINT # (B)(4). DATE SENT TO THE FDA: 10/14/2025. H6 COMPONENT CODE: G07002 - PENDING EVALUATION OF RETURNED DEVICE. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT: WHEN DID THE NEEDLES DETACH FROM THE SUTURE THREAD (REMOVAL FROM PACKAGE / DURING HANDLING PRIOR TO USE ON PATIENT/ DURING PASSAGE THROUGH TISSUE / DURING TYING / POST-OP)? HOW MANY NEEDLES DETACH FROM THE SUTURE THREAD? PROVIDE THE SOURCE OR NAME AND TITLE OF EXTERNAL PERSON PROVIDING ANSWERS TO FOLLOW-UP (EXTERNAL PERSON SUBMITTING ANSWERS TO SALES REP). A DEVICE HAS BEEN RECEIVED, HOWEVER IT HAS NOT YET BEEN EVALUATED. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
Additional Manufacturer Narrative
PRODUCT COMPLAINT #: (B)(4). H6 COMPONENT CODE: G07002 - NO DEVICE PROBLEM FOUND. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: H6, D9, H3. H3 INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO ETHICON FOR EVALUATION. EIGHT REPRESENTATIVE UNOPENED SAMPLES WERE RECEIVED FOR ANALYSIS. PRODUCT CODE M660G. THE COMPLAINT SAMPLE WAS NOT RECEIVED FOR EVALUATION. UPON INITIAL INSPECTION OF THE SAMPLES, NO EXTERNAL DAMAGES WERE OBSERVED ON THE EXTERNAL PACKETS. IN ORDER TO EVALUATE THE CONDITIONS OF THE RETURNED SAMPLES, THE PACKETS WERE OPENED, AND NO DEFECTS WERE DETECTED. THE SWAGE AND ATTACHMENT AREA WERE NOTED TO BE AS EXPECTED. THE SUTURES WERE DISPENSED WITHOUT PROBLEMS AND EXAMINED ALONG THE STRAND TO DETECT ANY ISSUE RELATED NO DEFECTS WERE OBSERVED DURING EVALUATION. A FUNCTIONAL TEST WAS PERFORMED USING INSTRON EQUIPMENT, AND THE PULL FORCE RESULTS WERE ABOVE THE MINIMUM REQUIREMENTS. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DEFECT NOTED. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH BATCH IS RANDOMLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE EVENT DESCRIBED COULD NOT BE CONFIRMED AS NO PRODUCT DEFECT WAS IDENTIFIED, THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT COULD NOT BE REPLICATED DURING THE LABORATORY ANALYSIS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE 100XGD / M660G BATCH NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED.