inforMED
InjuryITJ

DRIVE NITRO SPRINT ROLLATOR

Received Oct 15, 2025

Report 2438477-2025-00055 · MDR key 23280158

Device

Generic name

Walker, Mechanical

Model number

102662BK-T

Product problems

  • Mechanical Problem
  • Mechanical Problem

Patient

58 YR · Unknown

  • Bone Fracture(s)
  • Bone Fracture(s)

Narrative

Description of Event or Problem

DRIVE DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A ROLLATOR BY THE END USER WHO STATED THAT HE WAS STANDING WITH THE UNIT WITH THE BRAKES LOCKED AND THE BRAKES SUDDENLY RELEASED CAUSING THE ROLLATOR TO SPIN, AND HE FELL TO THE GROUND RESULTING IN A BROKEN RIB. AS PART OF ITS COMPLAINT REVIEW AND EVALUATION PROCESS, DRIVE DEVILBISS HEALTHCARE WILL ALSO NOTIFY THE MANUFACTURER OF THE PRODUCT ABOUT THIS COMPLAINT.

Additional Manufacturer Narrative

DRIVE DEVILBISS HEALTHCARE RECEIVED THE DEVICE RETURNED FOR EVALUATION. THE BRAKES WERE TESTED IN ALL POSITIONS AND FUNCTIONED AS DESIGNED.