MalfunctionFLE

AMSCO EVOLUTION

Received Oct 10, 2025 · Event occurred Sep 10, 2025

Report 3005899764-2025-00083 · MDR key 23270905

Device

Generic name

Amsco Evolution Sterilizer

Model number

HC-1500

Catalog number

UE06

Product problems

  • Detachment of Device or Device Component
  • Detachment of Device or Device Component

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER FACILITY REPORTED THAT THE UPPER PANEL OF THEIR AMSCO EVOLUTION STERILIZER FELL AND CONTACTED AN EMPLOYEE'S HEAD. THE EMPLOYEE DID NOT SEEK MEDICAL TREATMENT AND CONTINUED TO WORK. NO REPORT OF INJURY.

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE UNIT AND FOUND THAT THE SCREWS THAT HOLD THE UPPER PANEL WERE LOOSE, SUBSEQUENTLY CAUSING THE REPORTED EVENT. A CAUSE OF THE LOOSE SCREWS COULD NOT BE DETERMINED. TO RESOLVE THE ISSUE, THE TECHNICIAN REATTACHED THE UPPER PANEL AND THE SCREWS THAT HOLD IT. THE UNIT WAS TESTED, CONFIRMED TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED TO SERVICE. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.