inforMED
MalfunctionFLE

AMSCO

Received Oct 10, 2025 · Event occurred Sep 11, 2025

Report 3005899764-2025-00084 · MDR key 23268650

Device

Generic name

Amsco 600 Sterilizer

Catalog number

EF1101122

Product problems

  • Leak/Splash
  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE TO INSPECT THE AMSCO 600 STERILIZER AND FOUND THAT THE HEATING ELEMENT OF THE UNIT WAS DAMAGED RESULTING IN THE REPORTED EVENT. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED. THE TECHNICIAN REPLACED THE HEATING ELEMENT AND PROACTIVELY REPLACED THE CHECK VALVE, SAFETY VALVE, GENERATOR CONTACTOR AND WATER PUMP. THE UNIT WAS TESTED, CONFIRMED TO BE OPERATING ACCORDING TO SPECIFICATION, AND RETURNED TO SERVICE. A 3-YEAR COMPLAINT REVIEW CONFIRMED THIS TO BE AN ISOLATED EVENT. NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT THEIR AMSCO 600 STERILIZER LEAKED STEAM AND WATER ONTO THE FLOOR. NO REPORT OF INJURY.