MEDEX EXTENSION FILTER SETS
Received Oct 8, 2025 · Event occurred Aug 11, 2025
Report 1526863-2025-00120 · MDR key 23245857
Device
Generic name
Filter, Infusion Line
Manufacturer
Icu Medical Healthcare Manufacturing S.a. De C.vCatalog number
MX448HFB
Lot number
4425016
Product problems
- Contamination
- Contamination
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
Description of Event or Problem
IT WAS REPORTED THAT THE DEVICE EXHIBITED A BLACK DOT IN THE WHITE CONNECTOR. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.
Additional Manufacturer Narrative
TWO UNUSED DEVICES FROM THE SAME LOT WERE RETURNED FOR INVESTIGATION. THE DEVICES WERE VISUALLY INSPECTED AND THERE WAS NO CONTAMINATION WAS OBSERVED ON COMPONENT. DURING THE PHOTO EVALUATION, THE CUSTOMER REPORTED ISSUE WAS CONFIRMED SINCE AN EMBEDDED BLACK DOT ON COMPONENT. HOWEVER, CONDITION REPORTED WAS NOT DETECTED THROUGH THE ANALYSIS OF THE RETURNED SAMPLES. THE CAUSE FOR THIS FAILURE MODE IS RELATED TO A CONTAMINATION DURING THE MOLDING PROCESS OF THIS COMPONENT. A DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED WHICH INDICATED ALL INSPECTIONS WERE COMPLETED AND NO ISSUES WERE NOTED DURING MANUFACTURE.