ARCHITECT FREE T4
Received Oct 7, 2025 · Event occurred Sep 30, 2025
Report 3005094123-2025-00506 · MDR key 23238202
Device
Generic name
Radioimmunoassay, Free Thyroxine
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07K65-29
Lot number
73547UD00
Product problems
- High Test Results
- High Test Results
Patient
37 YR · Male
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, A TICKET TRENDING REVIEW, A DEVICE HISTORY RECORD REVIEW, AND A LABELING REVIEW. IN ADDITION, IN-HOUSE TESTING OF RETAINED REAGENT KIT WAS ALSO COMPLETED. THE DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. A REVIEW OF TRACKING AND TRENDING DATA FOR THE ARCHITECT FREE T4 ASSAY (LIST NUMBER 07K65) DID IDENTIFY AN INCREASE IN COMPLAINTS. HOWEVER, AN INCREASE IN COMPLAINT ACTIVITY FOR LOT 73547UD00 WAS NOT IDENTIFIED. ADDITIONALLY, IN-HOUSE PERFORMANCE TESTING WAS PERFORMED UTILIZING RETAINED REAGENT KIT OF LOT 73547UD00. ALL TESTING PASSED INDICATING THE REAGENT LOT IS PERFORMING AS EXPECTED. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS WITH LOT NUMBER 73547UD00 AND THE COMPLAINT ISSUE. A REVIEW OF LABELING WAS PERFORMED AND FOUND TO SUFFICIENTLY ADDRESS THE CUSTOMER'S ISSUE. BASED ON THIS INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED WITH THE ARCHITECT FREE T4 REAGENT, LOT NUMBER 73547UD00.
Description of Event or Problem
THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT FREE T4 RESULT FOR ONE 37-YEAR-OLD MALE PATIENT. THE SAMPLE WAS REPEATED WITH LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6), PROCESSED ON (B)(6) 2025 INITIAL RESULT = >64.35 PMOL/L REPEATED ON (B)(6) 2025 = 12.2/12.2/12.7 NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
Description of Event or Problem
THE CUSTOMER OBSERVED A FALSELY ELEVATED ARCHITECT FREE T4 RESULT FOR ONE 37-YEAR-OLD MALE PATIENT. THE SAMPLE WAS REPEATED WITH LOWER RESULTS. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6) PROCESSED ON (B)(6) 2025 INITIAL RESULT = >64.35 PMOL/L REPEATED ON (B)(6) 2025 = (B)(6) NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.