PORTEX TUBES BLUSELECT SUCTIONAID
Received Oct 6, 2025 · Event occurred Sep 2, 2025
Report 3012307300-2025-11287 · MDR key 23229568
Device
Generic name
Bag, Reservoir
Manufacturer
Icu Medical Czech Republic A. SCatalog number
101/870/070CZ
Lot number
NI
Product problems
- Crack
- Crack
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H6: CODES WERE UPDATED. ONE DEVICE WAS RETURNED FOR INVESTIGATION. IT WAS REPORTED THAT VISUAL INSPECTION SUCTION OF CONNECTOR, NO PHYSICAL DAMAGE OR OTHER ANOMALIES OBSERVED. WHEN ACCESSED WITH A LUER SLIP SYRINGE, A CRACK WAS OBSERVED. THE COMPLAINT OF SUCTION PORT BEING CRACKED CAN BE CONFIRMED. THE PROBABLE CAUSE OF THE CRACK IS UNKNOWN. A DEVICE HISTORY RECORD (DHR) REVIEW REPORTED NO DISCREPANCIES OR NON-CONFORMANCES DURING THE MANUFACTURING OF THE REPORTED LOT NUMBER.
Description of Event or Problem
IT WAS REPORTED THAT THE DEVICE WAS FOUND CRACKED AFTER 4 DAYS OF INSERTION. THE EVENT OCCURRED DURING USE ON PATIENT.
Additional Manufacturer Narrative
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER.