ELECSYS CMV IGM
Received Oct 6, 2025 · Event occurred Jul 29, 2025
Report 1823260-2025-04080 · MDR key 23227165
Device
Generic name
Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus
Manufacturer
Roche DiagnosticsCatalog number
07027133190
Lot number
84802901
Product problems
- False Negative Result
- Low Test Results
- False Negative Result
- Low Test Results
Patient
34 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE ANALYZER SERIAL NUMBER IS (B)(6). THE CALIBRATION AND QC RECOVERY DATA PROVIDED WAS ACCEPTABLE. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS CMV IGM AND ELECSYS CMV IGG RESULTS FOR 1 PATIENT SAMPLE ON A COBAS E 801 ANALYTICAL UNIT COMPARED TO TWO COMPETITOR METHODS. THIS MEDWATCH WILL COVER CMV IGM. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE CMV IGG RESULTS. THE INITIAL IGM RESULT WAS 0.22 COI, INTERPRETED AS NON-REACTIVE. THE INITIAL IGG RESULT WAS< 0.25 U/ML, INTERPRETED AS NON-REACTIVE. THE SAMPLE WAS TESTED ON A DIASORIN LIASON ANALYZER AND THE IGM RESULT WAS NON-REACTIVE, THE IGG RESULT WAS 44.5, INTERPRETED AS REACTIVE, AND THE IGG AVIDITY RESULT WAS HIGH AVIDITY. THE IGG RESULT UNITS OF MEASUREMENT WERE NOT PROVIDED. THE SAMPLE WAS ALSO TESTED ON A VIDAS BIOMÉRIEUX ANALYZER AND THE IGM RESULT WAS 3.12, INTERPRETED AS REACTIVE, THE IGG RESULT WAS 12, INTERPRETED AS REACTIVE, AND THE IGG AVIDITY RESULT WAS INTERMEDIATE AVIDITY. THE IGM AND IGG RESULTS UNITS OF MEASUREMENT WERE NOT PROVIDED.
Additional Manufacturer Narrative
NO FURTHER INFORMATION WAS PROVIDED FROM THE CUSTOMER SITE. SAMPLE MATERIAL WAS REQUESTED BUT WAS NOT PROVIDED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED.