inforMED
MalfunctionJIQ

TOTAL PROTEIN URINE/CSF GEN.3

Received Oct 6, 2025 · Event occurred Sep 12, 2025

Report 1823260-2025-04075 · MDR key 23223927

Device

Generic name

Urine/csf Total Protein Test

Manufacturer

Roche Diagnostics

Catalog number

08058679190

Lot number

852901

Product problems

  • False Positive Result
  • Incorrect Measurement
  • False Positive Result
  • Incorrect Measurement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE ANALYZER SERIAL NUMBER IS (B)(6). THE CUSTOMER SUSPECTS AN INTERFERING FACTOR IN THE PATIENT SAMPLE. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE TOTAL PROTEIN URINE/CSF GEN.3 RESULTS FOR 1 PATIENT URINE SAMPLE ON A COBAS C 503 ANALYTICAL UNIT. THE CUSTOMER QUESTIONED THE INITIAL RESULT AS A NEGATIVE PROTEIN RESULT WAS EXPECTED. THE INITIAL RESULT WAS 93.8 MG/DL. THE SAMPLE WAS REPEATED ON ANOTHER C503 ANALYZER IN ANOTHER LABORATORY AND THE RESULT WAS 93.8 MG/DL. THE SAMPLE WAS REPEATED AGAIN ON A U601 ANALYZER AND THE RESULT WAS NEGATIVE.

Additional Manufacturer Narrative

SECTION A2 AND A3 WERE UPDATED, AS WELL AS PATIENT MEDICATIONS. CALIBRATION WAS LAST PERFORMED ON (B)(6) 2025. THE QC RECOVERY DATA PROVIDED WAS WITHIN SPECIFICATION. BASED ON THE CALIBRATION AND QC DATA, A GENERAL REAGENT ISSUE COULD BE EXCLUDED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE ROOT CAUSE OF THE EVENT WAS FOUND TO BE CONSISTENT WITH A PATIENT SAMPLE ISSUE.