Received Sep 26, 2025 · Event occurred Sep 5, 2025
Report 3007215625-2025-00700 · MDR key 23159681
Device
Generic name
Brush/dermabrasion/powered
Manufacturer
Allergan PleasantonCatalog number
A1002
Lot number
NI
Product problems
- Material Frayed
- Material Frayed
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A PHOTOGRAPH WAS PROVIDED. PER THE IMAGE PROVIDED THE INTERNAL WIRES ARE EXPOSED. A REPLACEMENT POWER CORD / SUPPLY RESOLVED THE ISSUE.
Description of Event or Problem
HEALTHCARE PROFESSIONAL REPORTED, "POWER CORD SPARKED AND NO LONGER PROVIDES POWER TO THE UNIT" FOR THE DERMALINFUSION SYSTEM. ISSUE WAS RESOLVED BY REPLACING POWER CORD / SUPPLY. THERE WAS NO PATIENT INVOLVEMENT.