inforMED
MalfunctionGFE

D-PRO

Received Sep 26, 2025 · Event occurred Sep 5, 2025

Report 3007215625-2025-00700 · MDR key 23159681

Device

Generic name

Brush/dermabrasion/powered

Catalog number

A1002

Lot number

NI

Product problems

  • Material Frayed
  • Material Frayed

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A PHOTOGRAPH WAS PROVIDED. PER THE IMAGE PROVIDED THE INTERNAL WIRES ARE EXPOSED. A REPLACEMENT POWER CORD / SUPPLY RESOLVED THE ISSUE.

Description of Event or Problem

HEALTHCARE PROFESSIONAL REPORTED, "POWER CORD SPARKED AND NO LONGER PROVIDES POWER TO THE UNIT" FOR THE DERMALINFUSION SYSTEM. ISSUE WAS RESOLVED BY REPLACING POWER CORD / SUPPLY. THERE WAS NO PATIENT INVOLVEMENT.