inforMED
MalfunctionCAL

PENTAX

Received Sep 25, 2025 · Event occurred Sep 17, 2025

Report 9610877-2025-53407 · MDR key 23140363

Device

Generic name

Fiberopticintubationscope 2.4tp Ro Pls

Model number

FP-7RBS2

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

G4: THIS DEVICE IS NOT DISTRIBUTED IN US SO THAT 510K# IS BLANK. D4 UDI: "NOT DISTRIBUTED IN USA", BECAUSE PRODUCTS ARE NOT DISTRIBUTED IN USA PRODUCTS, BUT SIMILAR DEVICE PRODUCT ARE LISTED IN USA. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CFB (COHERENT FIBER BUNDLE) BROKEN. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE CFB (COHERENT FIBER BUNDLE). IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE BENDING RUBBER PERFORATED, THE INSERTION FLEXIBLE TUBE CRUSHED, AND THE SEGMENT CRUSHED; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem

THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. FIBER IMAGE FAILURE (BROKEN).