APTIMA COMBO 2 ASSAY

Received Sep 24, 2025 · Event occurred Sep 22, 2025

Report 2024800-2025-00062 · MDR key 23138036

Device

Generic name

Dna-reagents, Neisseria

Manufacturer

Hologic, Inc.

Catalog number

ASY-15647

Lot number

922686

Product problems

  • Off-Label Use
  • Off-Label Use

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ON (B)(6) 2025, A CUSTOMER REPORTED TO HOLOGIC A DISCREPANT RESULT FOR SAMPLE ID (SID) (B)(6) WHILE USING APTIMA COMBO 2 ASSAY MASTER LOT 922686 ON THEIR PANTHER INSTRUMENT. ON (B)(6) 2025, CUSTOMER INITIALLY TESTED SID (B)(6) ON THE PANTHER PLUS INSTRUMENT SN (B)(6) ON WORKLIST (WL) (B)(4) AND RECEIVED A DUAL CT POSITIVE, GC POSITIVE RESULT. PER CUSTOMER LAB POLICY, A SAMPLE IS RETESTED IF THERE IS A DUAL CT POSITIVE, GC POSITIVE RESULT. ON (B)(6) 2025, CUSTOMER RETESTED THE SID (B)(6) ON (B)(4) USING APTIMA COMBO 2 ASSAY MASTER LOT 924264 ON A DIFFERENT PANTHER INSTRUMENT SN (B)(6) AND RECEIVED A DUAL CT NEGATIVE, GC NEGATIVE RESULT. CUSTOMER CONFIRMED THE DUAL CT NEGATIVE, GC NEGATIVE RESULT WAS REPORTED OUT. HOLOGIC HAS NOT BEEN INFORMED OF ANY PATIENT TREATMENT OR ANY ADVERSE PATIENT OUTCOMES RELATED TO THIS EVENT.

Additional Manufacturer Narrative

HOLOGIC REVIEWED THE PANTHER LOGS AND WORKLISTS AND NOTED THERE WERE NO INDICATIONS OF HARDWARE OR REAGENT PREPARATION ISSUES; THE MOST LIKELY CAUSE OF THE DISCREPANT RESULT WAS SAMPLE RELATED. HOLOGIC PERFORMED A RISK ASSESSMENT AND NOTED NO PRODUCT IMPACT. HOLOGIC HAS NOT BEEN INFORMED OF ANY ADVERSE PATIENT OUTCOMES RELATED TO THIS ISSUE.