inforMED
MalfunctionQJZ

XLUNG KIT 230

Received Sep 23, 2025 · Event occurred Aug 30, 2025

Report 3012172416-2025-00097 · MDR key 23127293

Device

Generic name

Extracorporeal System For Long-term Respiratory / Cardiopulmonary Failure

Manufacturer

Xenios Ag

Catalog number

32000014

Lot number

GSXA0602

Product problems

  • Contamination
  • Contamination /Decontamination Problem
  • Device Contaminated During Manufacture or Shipping
  • Contamination
  • Contamination /Decontamination Problem
  • Device Contaminated During Manufacture or Shipping

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: B5, D4, D9 PLANT INVESTIGATION: NON-CONFORMITY REPORTED DURING PRODUCTION DID NOT FOLLOW THE SAME FAILURE PATTERN. NO OTHER SIMILAR COMPLAINTS GAVE BEEN REPORTED FOR THIS BATCH NUMBER. THE COMPLAINT SAMPLE WAS RECEIVED ON SEPTEMBER 19, 2025. THE DESCRIBED BLACK DOTS AND AN ADDITIONAL RED MARK WERE FOUND ON THE TUBE IN THE TABLE TRAY. THESE WERE FOUND TO BE ON THE OUTSIDE OF THE TUBE. THE BLACK SPOTS COULD BE WIPED AWAY WITH A PAPER TOWEL SOAKED IN 70% ISOPROPYL ALCOHOL. THE RED MARK WAS NOT REMOVABLE AND MAY HAVE BEEN MADE WITH A PERMANENT MARKER. THE CAUSE WAS TRACED TO MANUFACTURING AND EMPLOYEES WERE MADE AWARE OF THIS ERROR PATTERN.

Description of Event or Problem

A USER FACILITY REPORTED THAT THEY DETECTED AT LEAST 5 BLACK DOTS ON THE PRODUCT, INDICATING CONTAMINATION IN THE STERILE PACKAGING OF THE XLUNG KIT. GIVEN THAT THIS UNIT COULD NOT BE DISASSEMBLED THE USER COULD NOT DETERMINE WHETHER THE CONTAMINATION IS ON THE OUTSIDE OR INSIDE OF THE TUBING SYSTEM, AND THEREFORE THE PRODUCT COULD NOT BE USED. THE USER DETECTED THE PROBLEM WHILE THE PRIMING FLUID WAS CIRCULATING, SO A MINIMAL AMOUNT OF PRIMING FLUID REMAINED IN THE SYSTEM. THERE WAS NO PATIENT INVOLVEMENT. THE COMPLAINT SAMPLE WAS RECEIVED BY XENIOS FOR PRODUCT INVESTIGATION.

Additional Manufacturer Narrative

THE PLANT INVESTIGATION IS IN PROCESS. A SUPPLEMENTAL MDR WILL BE SUBMITTED UPON COMPLETION OF THIS ACTIVITY.

Description of Event or Problem

A USER FACILITY REPORTED THAT THEY DETECTED AT LEAST 5 BLACK DOTS ON THE PRODUCT, INDICATING CONTAMINATION IN THE STERILE PACKAGING OF THE XLUNG KIT. GIVEN THAT THIS UNIT COULD NOT BE DISASSEMBLED THE USER COULD NOT DETERMINE WHETHER THE CONTAMINATION IS ON THE OUTSIDE OR INSIDE OF THE TUBING SYSTEM, AND THEREFORE THE PRODUCT COULD NOT BE USED. THE USER DETECTED THE PROBLEM WHILE THE PRIMING FLUID WAS CIRCULATING, SO A MINIMAL AMOUNT OF PRIMING FLUID REMAINED IN THE SYSTEM. THERE WAS NO PATIENT INVOLVEMENT. THE COMPLAINT SAMPLE WAS AVAILABLE FOR PRODUCT INVESTIGATION.