inforMED
MalfunctionJCF

BD VACUTAINER® CPT® NC: 1.0ML FICOLL¿: 2.0ML

Received Sep 23, 2025 · Event occurred Apr 24, 2025

Report 1917413-2025-01341 · MDR key 23127238

Device

Generic name

Lymphocyte Separation Medium

Catalog number

362782

Lot number

4346579

Product problems

  • Break
  • Break

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

REPORT 3 OF 4: IT WAS REPORTED WHILE USING BD VACUTAINER® CPT® NC: 1.0ML FICOLL¿: 2.0ML, 1 TUBE BROKE DURING CENTRIFUGATION. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.

Additional Manufacturer Narrative

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 60 RETENTION SAMPLES FROM BOTH LOTS WERE INSPECTED AND NO ISSUES OR CRACKS WERE OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. A COMPLAINT HISTORY WAS PERFORMED AND NO TREND WAS OBSERVED. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE GLASS BREAKAGE DURING CENTRIFUGATION. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.