SODIUM ELECTRODE
Received Sep 23, 2025 · Event occurred Sep 3, 2025
Report 1823260-2025-03451 · MDR key 23124328
Device
Generic name
Electrode, Ion Specific, Sodium
Manufacturer
Roche DiagnosticsCatalog number
10825468001
Lot number
EHJ10
Product problems
- High Test Results
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE SERIAL NUMBER OF THE C 501 ANALYZER IS (B)(6). THE FIELD SERVICE ENGINEER DETERMINED THE SODIUM SENSOR WAS CAUSING AIR IN THE FLOW PATH. THE ISE REAGENTS, SIPPER PROBE, TUBING, AND SODIUM SENSOR WERE REPLACED. THE ISE MEASUREMENT BLOCK WAS CLEANED. THE SYSTEM REAGENTS WERE PRIMED AND THE ISE FLOW PATH WAS PROTEINIZED. ISE CHECKS, CONTROLS, AND PRECISION STUDIES WERE RUN. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO PATIENT SAMPLES TESTED WITH THE SODIUM ELECTRODE ON A COBAS 6000 C (501) MODULE. THE CUSTOMER REPEATED PATIENT SAMPLES SINCE THEY WERE TROUBLESHOOTING ISSUES WITH CALIBRATION.. THE FIRST SAMPLE INITIALLY RESULTED IN A SODIUM VALUE OF 146 MMOL/L AND IT REPEATED AS 141 MMOL/L. THE SAMPLE WAS REPEATED ON A SECOND ANALYZER, RESULTING IN A VALUE OF 138 MMOL/L. THE VALUE OF 138 MMOL/L WAS DEEMED CORRECT. THE SECOND SAMPLE INITIALLY RESULTED IN A SODIUM VALUE OF 136 MMOL/L AND IT REPEATED AS 131 MMOL/L. THE SAMPLE WAS REPEATED ON A SECOND ANALYZER, RESULTING IN A VALUE OF 130 MMOL/L.
Additional Manufacturer Narrative
CALIBRATION DATA WAS ACCEPTABLE. THE INVESTIGATION DETERMINED THE SERVICE ACTIONS RESOLVED THE ISSUE.