inforMED
MalfunctionJOL

J-VAC*RESV. 150 ML

Received Sep 23, 2025

Report 2210968-2025-10874 · MDR key 23123281

Device

Generic name

Catheter And Tip Suction

Manufacturer

Ethicon Inc.

Catalog number

2177

Lot number

JU0852

Product problems

  • Obstruction of Flow
  • Obstruction of Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON INC, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON INC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. ADDITIONAL INFORMATION: D9, H3, H4, H6 THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WAS MET PRIOR TO THE RELEASE OF THIS LOT. ADDITIONAL H6 COMPONENT CODE: G07002 NO DEVICE PROBLEM ADDITIONAL H3 INVESTIGATION SUMMARY: MAN TIE TOO LONG DURING SAMPLE EVALUATION THE PLATE WAS ABLE TO BE OPEN AND CLOSE WITHOUT ANY ISSUE. THE STRAP DID NOT INTERFERE ON THE CORRECT FUNCTION OF THE LOCKING MECHANISM. THEREFORE, IS CONSIDERED THIS MAN FACTOR DOES NOT CONTRIBUTE TO THE REPORTED COMPLAINT DEFECT. MATERIALS DAMAGED/ BROKEN COMPONENT LOCKING MECHANISM OF RESERVOIR WAS CHECKED TO VERIFY ITS CONDITION. SAMPLE WAS EVALUATED, AND DURING THIS EVALUATION IT WAS NOTICE THAT THE LATCH OF PLATE AND ITS MECHANISM WERE IN GOOD CONDITIONS; THE PLATE WAS ABLE TO BE OPEN AND CLOSE. IN ADDITION, PLATE SUBASSEMBLY NEEDS TO BE PROPERLY CLOSED IN ORDER TO PERFORM FINAL ASSEMBLY ON FINISH GOOD. THEREFORE, IS CONSIDERED THIS MATERIALS FACTOR DOES NOT CONTRIBUTE TO THE REPORTED COMPLAINT DEFECT. BASED ON THE PRESENT ANALYSIS, AND CONDITION OF SAMPLE RECEIVED IT IS DETERMINED THAT THE REPORTED COMPLAINT IS NOT CONFIRMED AND IT IS NOT RELATED TO THE MANUFACTURING PROCESS AND OTHER EXTERNAL CAUSES COULD HAVE AFFECTED THE PRODUCT INTEGRITY SUCH AS HANDLING, NON-PROPER USAGE OR ANY OTHER POSSIBLE CONTRIBUTOR OUT OF THE MANUFACTURE CONTROL.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). ADDITIONAL INFORMATION: H6 COMPONENT CODE: G07002 - DEVICE NOT RETURNED. ADDITIONAL INFORMATION PROVIDED: THE RESERVOIR DID NOT SWELL EVEN WHEN BENDING THE BOTTOM FLAP. ADDITIONAL INFORMATION HAS BEEN REQUESTED AND RECEIVED. ATTEMPTS TO OBTAIN THE DEVICE HAVE BEEN MADE. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. WHAT IS THE LOT NUMBER? JU0852. PLEASE PERFORM AND DOCUMENT THE FOLLOW UP ATTEMPT FOR PRODUCT RETURN. WE REGULARLY CONTACT WITH SALE REP ABOUT THE DEVICE RETURNING. PLEASE PROVIDE THE SOURCE OR NAME AND TITLE OF EXTERNAL PERSON PROVIDING ANSWERS TO FOLLOW-UP (EXTERNAL PERSON SUBMITTING ANSWERS TO SALES REP OR LOC/BQ) (B)(6). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY ETHICON, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, ETHICON, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

IT WAS REPORTED A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND A RESERVOIR WAS USED. THE RESERVOIR COULD NOT BE LOCKED. THE DETAILS ARE UNKNOWN AT THIS TIME, SO IT IS BEING CONFIRMED. ANOTHER PRODUCT WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. FURTHER DETAILS ARE NOT PROVIDED. ADDITIONAL INFORMATION HAS BEEN REQUESTED.