inforMED
MalfunctionPOL

TREVO NXT 6MM X 37MM - US

Received Sep 22, 2025 · Event occurred Aug 30, 2025

Report 3012931345-2025-00155 · MDR key 23111218

Device

Generic name

Neurovascular Mechanical Thrombectomy Device For Acute Ischemic Stroke Treatment

Catalog number

90315

Lot number

0000774056

Product problems

  • Fracture
  • Fracture

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THERE ARE CONTROLS IN THE MANUFACTURING PROCESS TO ENSURE THE PRODUCT MET SPECIFICATIONS UPON RELEASE. DURING VISUAL INSPECTION, THE DISTAL END OF THE RETRIEVER CORE WIRE WAS OBSERVED TO HAVE BEEN FRACTURED AT 5.5CM FROM THE DISTAL END OF THE CORE WIRE. THERE WAS EVIDENCE OF STRETCHING TO THE PEBAX/LAMINATION AND EVIDENCE OF NECKING/STRETCHING TO THE CORE WIRE FRACTURE POINT. THE RETRIEVER SHAPED SECTION WAS INTACT WITH NO DAMAGE NOTED. A FUNCTIONAL TEST COULD NOT BE PERFORMED AS THE RETRIEVER CORE WIRE WAS FRACTURED. THE REPORTED EVENT OF ¿RETRIEVER CORE BROKEN DURING USE¿ WAS CONFIRMED DURING THE DEVICE ANALYSIS. THE REPORTED EVENT OF ¿RETRIEVER DIFFICULT/UNABLE TO GO THROUGH THE CATHETER SHAFT¿ WAS NOT REPLICATED, AS A FUNCTIONAL TEST COULD NOT BE PERFORMED, HOWEVER THE ANALYSIS RESULTS ARE CONSISTENT WITH THE REPORTED EVENTS. THE DEVICE FAILED TO MEET SPECIFICATIONS WHEN RECEIVED, BASED ON THE DAMAGE NOTED. THE REPORTED EVENT IS COVERED IN THE DEVICE DIRECTIONS FOR USE (DFU). AS WELL, THE RISK OF THE REPORTED EVENT IS DOCUMENTED IN THE RISK DOCUMENTATION AND THERE ARE CURRENT CONTROLS TO MITIGATE THE RISK OF THE AS REPORTED EVENT. IT WAS REPORTED THAT ALTHOUGH FEELING STRONG RESISTANCE BETWEEN THE SUBJECT STENT RETRIEVER AND A MICROCATHETER, PROCEDURE WAS CONTINUED, AND THE THROMBUS WAS RETRIEVED USING AN ASPIRATION CATHETER. WHEN THE RETRIEVED THROMBUS WAS CONFIRMED THAT THE DELIVERY WIRE WHICH IS COVERED WITH THE POLYMER JACKET WAS FOUND TO BE FRACTURED. AS PER THE ADDITIONAL INFORMATION, FORCE WAS APPLIED AGAINST THE RESISTANCE, THE PRODUCT WAS PROPERLY SIZED FOR THE VESSEL DIAMETER, INTERMITTENT FLUSH WAS USED THROUGHOUT THE CLINICAL PROCEDURE. IT IS UNKNOWN IF THE PATIENTS ANATOMY WAS TORTUOUS. AS PER THE TREVO RETRIEVER IFU: TO PREVENT THROMBUS FORMATION AND CONTRAST MEDIA CRYSTAL FORMATION, MAINTAIN A CONSTANT INFUSION OF APPROPRIATE FLUSH SOLUTION THROUGH ALL CATHETER LUMENS. IMPORTANT: INTRODUCE INSERTION TOOL HALF WAY INTO THE ROTATING HEMOSTASIS VALVE AND USE SYRINGE OR INFUSION LINE TO FLUSH INSERTION TOOL UNTIL SALINE EXITS THE PROXIMAL END OF THE INSERTION TOOL. IF INSERTION TOOL IS NOT PROPERLY FLUSHED, IT MAY BE DIFFICULT TO ADVANCE THE RETRIEVER THROUGH THE INSERTION TOOL. DO NOT ADVANCE OR WITHDRAW RETRIEVER AGAINST RESISTANCE OR SIGNIFICANT VASOSPASM. MOVING OR TORQUING DEVICE AGAINST RESISTANCE OR SIGNIFICANT VASOSPASM MAY RESULT IN DAMAGE TO VESSEL OR DEVICE. ASSESS CAUSE OF RESISTANCE USING FLUOROSCOPY AND IF NEEDED RESHEATH THE DEVICE TO WITHDRAW. THE DEVICE WAS RETURNED, AND IT WAS CONFIRMED THAT THE RETRIEVER CORE WIRE HAD BEEN FRACTURED, THERE WAS ALSO DAMAGE NOTED TO THE PEBAX LAMINATION AT THE FRACTURE LOCATION. AS IT WAS STATED THAT INTERMITTENT FLUSH WAS USED DURING THE PROCEDURE, THIS IS THE LIKELY CAUSE FOR THE RESISTANCE EXPERIENCED DURING THE RETRIEVER ADVANCEMENT. THE RETRIEVER WAS CONTINUED TO BE ADVANCED AGAINST THE RESISTANCE WHICH IS LIKELY TO HAVE CAUSED THE FRACTURE AND DAMAGE NOTED TO THE RETRIEVER. BASED ON REVIEW OF ALL AVAILABLE INFORMATION AN ASSIGNABLE CAUSE OF USE ERROR WILL BE ASSIGNED TO THE REPORTED EVENT OF ¿RETRIEVER DIFFICULT/UNABLE TO GO THROUGH CATHETER SHAFT¿ AND ¿RETRIEVER CORE WIRE BROKEN DURING USE¿ AND TO THE ANALYSED EVENTS OF ¿RETRIEVER CORE WIRE BROKEN/FRACTURED DURING USE¿ AND ¿RETRIEVER DELIVERY WIRE LAMINATION ISSUE¿, AS THE ISSUE INVESTIGATION CONFIRMS THAT THERE WAS AN ACT OR OMISSION OF AN ACT THAT RESULTED IN A DIFFERENT MEDICAL PRODUCT RESPONSE THAN INTENDED BY THE MANUFACTURER AND/OR EXPECTED BY THE USER.

Description of Event or Problem

IT WAS REPORTED THAT THE STENT RETRIEVER (SUBJECT DEVICE) WAS USED IN A THROMBUS RETRIEVAL PROCEDURE. HOWEVER, RESISTANCE WAS ENCOUNTERED WHILE ADVANCING THE SUBJECT DEVICE WITHIN THE MICROCATHETER. AFTER THE THROMBUS WAS RETRIEVED THE DELIVERY WIRE OF THE SUBJECT DEVICE WAS OBSERVED TO BE FRACTURED, THE FRACTURED PART WAS WITHIN THE ASPIRATION CATHETER AND WAS RETRIEVED. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT.

Description of Event or Problem

IT WAS REPORTED THAT THE STENT RETRIEVER (SUBJECT DEVICE) WAS USED IN A THROMBUS RETRIEVAL PROCEDURE. HOWEVER, RESISTANCE WAS ENCOUNTERED WHILE ADVANCING THE SUBJECT DEVICE WITHIN THE MICROCATHETER. AFTER THE THROMBUS WAS RETRIEVED THE DELIVERY WIRE OF THE SUBJECT DEVICE WAS OBSERVED TO BE FRACTURED, THE FRACTURED PART WAS WITHIN THE ASPIRATION CATHETER AND WAS RETRIEVED. THE SUBJECT DEVICE WAS REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT.