CARDIOCHEK PLUS ANALYZER

Received Sep 21, 2025 · Event occurred Aug 18, 2025

Report 1836135-2025-00014 · MDR key 23106389

Device

Generic name

Clinical Chemistry Analyzer

Manufacturer

Pts Diagnostics

Model number

1.09

Product problems

  • Electrical /Electronic Property Problem
  • Electrical /Electronic Property Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THERE WERE NO ALLEGATIONS OF PATIENT OR USER HARM. THE CUSTOMER REPORTED THAT "WHILE TRYING TO TROUBLESHOOT, THE BATTERIES/BATTERY COMPARTMENT BECAME VERY HOT TO THE TOUCH."

Additional Manufacturer Narrative

CURRENTLY IT IS UNKNOWN WHETHER THE DEVICE MAY HAVE MALFUNCTIONED, AS THE ALLEGATION COULD NOT BE DUPLICATED. IT IS KNOWN THAT IMPROPER INSERTION OF BATTERIES - IN ANY DEVICE - MAY RESULT IN OVERHEATING, BATTERY LEAKAGE, ETC., WHICH IS STATED IN THE LABELING OF MOST BATTERY MANUFACTURERS. THE REPORTED DEVICE WAS BEING USED BEYOND ITS' RECOMMENDED DEVICE LIFE. THIS REPORT IS BEING SUBMITTED OUT OF AN ABUNDANCE OF CAUTION.