REFOBACIN BONE CEMENT R 10X2 GERMAN
Received Sep 17, 2025 · Event occurred Jul 3, 2025
Report 3006946279-2025-00107 · MDR key 23072367
Device
Generic name
Bone Cement
Manufacturer
Biomet France S.a.r.l.Catalog number
3009890001
Lot number
A335AC1403
Product problems
- Loosening of Implant Not Related to Bone-Ingrowth
- Loosening of Implant Not Related to Bone-Ingrowth
Patient
53 YR · Female
- Pain
- Loss of Range of Motion
- Aseptic loosening
- Pain
- Loss of Range of Motion
- Aseptic loosening
Narrative
Additional Manufacturer Narrative
(B)(4). D-4: THIS DEVICE IS SOLD OUTSIDE THE US AND THEREFORE NO GUDID INFORMATION EXISTS. THIS DEVICE IS CONSIDERED SIMILAR TO (B)(4). D6A - PRODUCT IMPLANTED IN 2014 OUTSIDE HUG. D10 ¿ (B)(6), OXFORD DOMED LAT MEN BRNG MD SZ 4, LOT 3203664. (B)(6), AGC DA 2000 FEMORAL LEFT 60MM, LOT 3194474. (B)(6), REFOBACIN BONE CEMENT R 10X2 GERMAN, LOT C308AC1403. (B)(6), OXF DOMED LAT TIB TRAY SZ C-UK, LOT 3349283. G2 ¿ FOREIGN ¿ SWITZERLAND. G-4: THE REPORTED PRODUCT IS NOT SOLD IN THE US, THE PRE-MARKET SUBMISSION NUMBER FOR THE SIMILAR PRODUCT SOLD IN THE US IS K171540. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT A REVISION SURGERY OF THE PARTIAL LEFT KNEE PROSTHESIS ELEVEN YEARS POST IMPLANTATION. DUE TO A FALSE MOVEMENT TWO MONTHS PRIOR REVISION, THE SUB-LUXATION OF THE POLYETHYLENE OCCURRED, AND THE PATIENT HAS BEEN IN PAIN, WITH PERSISTENT EXTENSION DEFICIT. RADIOLOGICAL EXAMINATION REVEALED LOOSENING OF THE DAMAGED TIBIAL COMPONENT DUE TO BLOCKAGE OF THE WORN MOBILE POLYETHYLENE BETWEEN THE NOTCH AND THE EXTERNAL JOINT SPACE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Additional Manufacturer Narrative
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, B5, G3, G6, H2, H3, H6, H10, H11. NO PRODUCT WAS RETURNED OR PICTURES PROVIDED; A PRODUCT EVALUATION COULD NOT BE PERFORMED. A REVIEW OF THE DEVICE MANUFACTURING RECORDS CONFIRMED NO ABNORMALITIES OR DEVIATIONS. REVIEW OF THE COMPLAINT HISTORY FOUND NO ADDITIONAL RELATED COMPLAINTS FOR THIS ITEM AND THE REPORTED PART AND LOT COMBINATION. THE REPORTED EVENT WAS CONFIRMED BY REVIEW OF X-RAYS. RADIOLOGICAL EXAMINATION REVEALS LOOSENING OF THE DAMAGED TIBIAL COMPONENT DUE TO BLOCKAGE OF THE WORN MOBILE POLYETHYLENE BETWEEN THE NOTCH AND THE EXTERNAL JOINT SPACE. THE DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
Description of Event or Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.