3M ESPE LAVA CHAIRSIDE ORAL SCANNER C.O.S.

Received Oct 19, 2011 · Event occurred Sep 19, 2011

Report 3005174370-2011-00011 · MDR key 2304763

Device

Generic name

Optical Impression System

Catalog number

68901

Product problems

  • Electrostatic Discharge

Patient

Not reported

  • Tingling
  • Electric Shock

Narrative

Description of Event or Problem

DR (B)(6) DENTAL OFFICE PURCHASED A NEW LAVA CHAIRSIDE ORAL SCANNER (C.O.S.). ON (B)(6) 2011, INTRODUCTION FOR THEORY AND PRACTICE TOOK PLACE USING AN ARTIFICIAL MODEL (PHANTOM HEAD). IN ADDITION, ONE DENTAL ASSISTANT AND TWO PATIENTS WERE PART OF THE TRAINING (THEY WERE SCANNED USING THE LAVA C.O.S.). THERE WERE NO COMPLAINTS ON THIS DAY (B)(6) ABOUT A STRANGE FEELING DURING SCANNING. ON THE (B)(6) 2011, A PATIENT REPORTED THAT SHE FELT A SHORT TINGLE SIMILAR TO A MILD ELECTRICAL SHOCK. THE TREATMENT OF THE PATIENT WAS STOPPED AFTER THE PATIENT TOLD THE DOCTOR ABOUT THE SENSATION. SUBSEQUENTLY, (B)(6) ONE OF THE ASSISTANTS ALSO REPORTED THAT SHE HAD A SIMILAR FEELING THE DAY BEFORE (B)(6). BOTH THE PATIENT AND THE ASSISTANT DESCRIBED THE EXPERIENCE AS A SHORT TINGLE. OTHER SYMPTOMS OR LONG-LASTING EFFECTS HAVE NOT BEEN REPORTED TO 3M ESPE.

Additional Manufacturer Narrative

RESULTS: THE DENTIST'S ELECTRICIAN WAS ON SITE ON (B)(6) 2011. THE ELECTRICIAN USED A SPECIAL MEASURING INSTRUMENT TO CHECK THE WHOLE ELECTRICAL INSTALLATION OF THE DENTAL OFFICE. THE MEASURING INSTRUMENT SHOWED THAT THE PROTECTIVE EARTH (GROUND) CONNECTION FROM THE POWER OUTLET WAS NOT WORKING. THE ELECTRICIAN FOUND THAT THE PROTECTIVE EARTH (GROUNDING) CONTACTS WERE OUT OF ORDER BECAUSE THEY WERE PAINTED WITH PAINT. ALL OF THE WALLS INSIDE THE DENTAL OFFICE HAD BEEN PAINTED IN 2011, DURING THAT TIME, THE PAINTER ACCIDENTALLY PAINTED OVER THE PROTECTIVE EARTH (GROUND) CONTACTS IN THE WALL OUTLETS. THE ELECTRICIAN USED SOME SANDING PAPER AND CLEANED THESE CONTACTS. AFTER HE HAD DONE THAT, HE MEASURED AGAIN THE COMPLETE INSTALLATION AND ALL VALUES WERE WITHIN REGULATIONS. CONCLUSION: AT THIS TIME 3M ESPE CONCLUDES THAT THE MILD ELECTRICAL SHOCK RECEIVED WAS CAUSED BY A LEAKAGE CURRENT OF THE LAVA C.O.S. SYSTEM, WHICH NORMALLY WOULD BE DISSIPATED BY A CORRECTLY GROUNDED OUTLET. THE 3M ESPE CONTINUES TO INVESTIGATE THIS EVENT AND WILL PROVIDE SUPPLEMENTAL FOLLOW-UP REPORTS, AS APPROPRIATE.