inforMED
InjuryKNP

DR COMFORT

Received Sep 12, 2025 · Event occurred Aug 18, 2025

Report 3008579854-2025-00002 · MDR key 23044768

Device

Generic name

Brian X Black 12 Med

Catalog number

6910-M-12.0

Lot number

UNKNOWN

Product problems

  • Patient Device Interaction Problem
  • Patient Device Interaction Problem

Patient

NA · Unknown

  • Ulcer
  • Ulcer

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A PLASTIC TOE PUFF THAT RUBBED THE PATIENTS HAMMER TOE WHICH LED TO AN ULCER AND THEN AMPUTATION.

Additional Manufacturer Narrative

IT WAS REPORTED THAT THERE WAS A PLASTIC TOE PUFF THAT RUBBED THE PATIENTS HAMMER TOE WHICH LED TO AN ULCER AND THEN AMPUTATION. THE SHOE WAS RETURNED AND EVALUATED. THE COMPLAINT HAS BEEN CONFIRMED; EVALUATION OF THE SHOE FOUND THE TOE BOX SUPPORT IS BENT DOWN AND CAN FEEL THE EDGE OF IT. IF ADDITIONAL INFORMATION REGARDING THE REPORTED EVENT IS SUBMITTED AT A FUTURE DATE, THIS INVESTIGATION WILL BE RE-EVALUATED.