inforMED
MalfunctionDXH

HOMECG+

Received Sep 12, 2025

Report 3012267350-2025-00013 · MDR key 23043096

Device

Generic name

Transmitters And Receivers, Electrocardiograph, Telephone

Model number

V2.0

Lot number

2021 - 16828

Product problems

  • Application Program Problem
  • Application Program Problem

Patient

NA · Male

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

THE USER REPORTS CHALLENGES WITH LOST CONTACT WHEN PLACING THUMBS ON THE ELECTRODES. THE USER HAD NO ADVERSE EVENT OR REACTION FROM THIS PROBLEM.

Additional Manufacturer Narrative

THE DEVICE APPEARED TO BE OPERATING ON AN OLDER FIRMWARE VERSION WHICH WAS CAUSING ISSUES WHEN ATTEMPTING TO TAKE SCANS. AFTER DELIVERY OF THE DEVICE WITH UPDATED FIRMWARE, THE USER WAS ABLE TO SUCCESSFULLY OBTAIN SCANS. ADDITIONALLY, THE USER RECEIVED WRIST ELECTRODES.

Remedial action

  • Repair