ALINITY I FREE T4 REAGENT KIT
Received Sep 12, 2025 · Event occurred Aug 24, 2025
Report 3005094123-2025-00453 · MDR key 23039282
Device
Generic name
Radioimmunoassay, Free Thyroxine
Manufacturer
Abbott Ireland Diagnostics DivisionCatalog number
07P70-20
Lot number
71236UD00
Product problems
- High Test Results
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF DATA AND INFORMATION PROVIDED BY THE CUSTOMER, A SEARCH FOR SIMILAR COMPLAINTS, A TICKET TRENDING REVIEW, A DEVICE HISTORY RECORD REVIEW, A LABELING REVIEW, AND IN-HOUSE TESTING OF RETAINED REAGENT KIT. THE DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. A SEARCH FOR SIMILAR COMPLAINTS IDENTIFIED AN INCREASE IN COMPLAINT ACTIVITY FOR LOT: 71236UD00. ADDITIONALLY, A REVIEW OF TRACKING AND TRENDING DATA FOR THE ALINITY I FREE T4 ASSAY IDENTIFIED AN INCREASE IN COMPLAINTS. HOWEVER, IN-HOUSE PERFORMANCE TESTING WAS PERFORMED UTILIZING RETAINED REAGENT KIT OF THE COMPLAINT LOT. ALL TESTING PASSED INDICATING THE REAGENT LOT IS PERFORMING AS EXPECTED. A REVIEW OF THE DEVICE HISTORY RECORD DID NOT IDENTIFY ANY NON-CONFORMANCE'S OR DEVIATIONS WITH LOT NUMBER: 71236UD00 AND THE COMPLAINT ISSUE. A REVIEW OF LABELING WAS PERFORMED AND FOUND TO SUFFICIENTLY ADDRESS THE CUSTOMER'S ISSUE. BASED ON THIS INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED WITH THE ALINITY I FREE T4 REAGENT, LOT NUMBER: 71236UD00.
Description of Event or Problem
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2025, SID: (B)(6): INITIAL RESULT = 33.41 PMOL/L, REPEAT = 13.39 PMOL/L. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Additional Manufacturer Narrative
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.
Description of Event or Problem
THE CUSTOMER REPORTED A FALSELY ELEVATED ALINITY I FREE T4 RESULT FOR ONE SAMPLE. THE FOLLOWING DATA WAS PROVIDED: ON (B)(6) 2025, SID (B)(6): INITIAL RESULT = 33.41 PMOL/L, REPEAT = 13.39 PMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.